用于心律失常诊断的无线心电图传感器系统的临床评估。

Clinical evaluation of a wireless ECG sensor system for arrhythmia diagnostic purposes.

机构信息

University of Agder, Faculty of Engineering and Science, Grimstad, Norway.

出版信息

Med Eng Phys. 2013 Jun;35(6):697-703. doi: 10.1016/j.medengphy.2013.03.002. Epub 2013 Apr 3.

Abstract

In a clinical study, a novel wireless electrocardiogram (ECG) recorder has been evaluated with regard to its ability to perform arrhythmia diagnostics. As the ECG recorder will detect a "non-standard" ECG signal, it has been necessary to compare those signals to "standard" ECG recording signals in order to evaluate the arrhythmia detection ability of the new system. Simultaneous recording of ECG signals from both the new wireless ECG recorder and a conventional Holter recorder was compared by two independent cardiology specialists with regard to signal quality for performing arrhythmia diagnosis. In addition, calculated R-R intervals from the two systems were correlated. A total number of 16 patients participated in the study. It can be considered that recorded ECG signals obtained from the wireless ECG system had an acceptable quality for arrhythmia diagnosis. Some of the patients used the wireless sensor while doing physical sport activities, and the quality of the recorded ECG signals made it possible to perform arrhythmia diagnostics even under such conditions. Consequently, this makes possible improvements in correlating arrhythmias to physical activities.

摘要

在一项临床研究中,评估了一种新型的无线心电图(ECG)记录器在进行心律失常诊断方面的能力。由于 ECG 记录器将检测到“非标准”ECG 信号,因此有必要将这些信号与“标准”ECG 记录信号进行比较,以评估新系统的心律失常检测能力。两名独立的心脏病专家比较了新的无线 ECG 记录器和传统的 Holter 记录器同时记录的 ECG 信号的质量,以便进行心律失常诊断。此外,还对两个系统的计算出的 R-R 间隔进行了相关性分析。共有 16 名患者参与了这项研究。可以认为,从无线 ECG 系统获得的记录 ECG 信号具有可接受的心律失常诊断质量。一些患者在进行体育活动时使用了无线传感器,并且记录的 ECG 信号的质量使得即使在这种情况下也能够进行心律失常诊断。因此,这使得将心律失常与身体活动相关联成为可能。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验