Sudden power loss in a HeartMate II left ventricular assist device due to intermittent pin contact with the battery: case report.

机构信息

Perfusion Services, University of Ottawa Heart Institute, Ottawa, Ontario, Canada.

出版信息

ASAIO J. 2013 May-Jun;59(3):317-9. doi: 10.1097/MAT.0b013e318288a3a8.

Abstract

Left ventricular assist device technology has improved such that mechanical malfunction, particularly with newer generation continuous flow devices, is a relatively rare event. We present a case of sudden power loss in a HeartMate II caused by intermittent contact of the battery terminals after a clip was dropped with the battery inserted in it. The clip was replaced and the patient made a complete recovery. A new inspection and testing methodology, and amended approach to patient and caregiver training, designed to prevent future occurrences is described.

摘要

左心室辅助装置技术已经得到了改进,使得机械故障,特别是新一代连续流装置的故障,变得相对罕见。我们报告了一例 HeartMate II 突然断电的病例,原因是电池插入后,夹子掉落在电池端子上,导致电池端子间歇性接触。更换了夹子,患者完全康复。描述了一种新的检查和测试方法,以及对患者和护理人员培训的改进方法,旨在防止未来再次发生这种情况。