Department of Pathology and Laboratory Medicine, University of Texas-Houston Medical School, Houston, TX 77030, USA.
Ther Drug Monit. 2013 Jun;35(3):384-7. doi: 10.1097/FTD.0b013e318288449a.
Phenytoin is a strongly protein-bound anticonvulsant requiring free drug monitoring for certain patients. Recently, free phenytoin assay on the TDx analyzer was discontinued by the Abbott Laboratories, but Roche Diagnostics has an assay for the determination of free phenytoin.
Because a clinical laboratory may not have an instrument for using Roche free phenytoin assay, we studied the possibility of determining free phenytoin using particle-enhanced turbidimetric inhibition immunoassay (PETINIA) total phenytoin assay on the Vista 1500 analyzer by comparing free phenytoin values obtained by Roche assay and Hitachi 917 analyzer with values obtained by using PETINIA assay.
When protein-free ultrafiltrates were prepared at room temperature from 41 different patients taking phenytoin, using Roche assay (x axis) and PETINIA (y axis) assay, the following regression equation was observed: y = 0.8526x + 0.0661 (r = 0.85, n = 41). The regression equation indicated significant negative bias with the reference method. However, when additional 30 specimens were studied where ultrafiltration was performed at 37°C, lesser negative bias was observed as revealed by the regression equation: y = 0.922x - 0.117 (r = 0.90, n = 30).
We conclude that PETINIA assay on the Vista 1500 analyzer is not suitable for monitoring free phenytoin concentration if ultrafiltration is performed at room temperature but may be used with caution if ultrafiltration is performed at physiological temperature of 37°C, which is the desirable temperature for preparation of ultrafiltrate.
苯妥英是一种与蛋白高度结合的抗惊厥药物,某些患者需要监测游离药物浓度。最近,雅培实验室停止了 TDx 分析仪上的游离苯妥英检测,但罗氏诊断公司有游离苯妥英检测方法。
由于临床实验室可能没有罗氏游离苯妥英检测方法所需的仪器,我们通过比较罗氏检测法和日立 917 分析仪与使用粒子增强浊度抑制免疫测定法(PETINIA)总苯妥英测定法获得的游离苯妥英值,研究了在 Vista 1500 分析仪上使用 PETINIA 测定法测定游离苯妥英的可能性。
当从 41 名服用苯妥英的不同患者中在室温下制备无蛋白超滤物时,使用罗氏检测法(x 轴)和 PETINIA (y 轴)检测法,观察到以下回归方程:y = 0.8526x + 0.0661(r = 0.85,n = 41)。回归方程表明与参考方法存在显著负偏倚。然而,当在 37°C 下进行超滤时,对另外 30 个标本进行研究,发现回归方程显示出较小的负偏倚:y = 0.922x - 0.117(r = 0.90,n = 30)。
我们的结论是,如果在室温下进行超滤,Vista 1500 分析仪上的 PETINIA 检测法不适合监测游离苯妥英浓度,但如果在生理温度 37°C 下进行超滤,可以谨慎使用,因为这是制备超滤物的理想温度。