Norris K A, Atkinson A R, Smith W G
Clin Chem. 1975 Jul;21(8):1093-101.
Phosphoenolpyruvate carbosylase (EC 4.1.1.31) has been used in developing a simple, inexpensive colorimetric assay for serum total carbon dioxide in an open system. The oxaloacetate formed by the action of the enzyme on bicarbonate and phosphoenolpyruvate is measured by use of the diazonium salt of Fast Violet B. Interferences from billirubin, pyruvate, and drugs are negligible. Acetoacetate interference is significant only in highly ketotic samples, and a serum blank corrects for it. Serum protein interference is equivalent to 3.3 plus or minus 1.25 mmol of CO2 per liter and hence is sufficiently constant to be corrected for by use of a serum standard or serum blank. The method has been applied to the Vickers M-300 and D-300 systems and within-batch standard deviations of plus or minus 0.1 to plus or minus 0.6 mmol/liter have been observed. Excellent correlation with orthodox techniques has been obtained.
磷酸烯醇丙酮酸羧化酶(EC 4.1.1.31)已被用于开发一种简单、廉价的比色法,用于在开放系统中测定血清总二氧化碳。通过使用固蓝B重氮盐来测量该酶作用于碳酸氢盐和磷酸烯醇丙酮酸所形成的草酰乙酸。胆红素、丙酮酸和药物的干扰可忽略不计。乙酰乙酸的干扰仅在高度酮症样本中显著,可通过血清空白进行校正。血清蛋白干扰相当于每升3.3±1.25 mmol二氧化碳,因此足够恒定,可通过使用血清标准品或血清空白进行校正。该方法已应用于维克斯M - 300和D - 300系统,观察到批内标准差为±0.1至±0.6 mmol/升。与传统技术具有良好的相关性。