Gagne Nolan L, Rivard Mark J
Department Radiation Oncology, Tufts University School of Medicine, Boston, MA 02111, USA.
Appl Radiat Isot. 2013 Sep;79:62-6. doi: 10.1016/j.apradiso.2013.04.034. Epub 2013 May 7.
This study explores the influence of source photon energy on eye plaque brachytherapy dose distributions for a 16 mm COMS plaque filled with (103)Pd, (125)I, or (131)Cs sources or monoenergetic photon emissions ranging from 12 keV to 100 keV. Dose distributions were similarly created for all permutations of three common brachytherapy seed designs. Within this range, sources with average energy ≤22 keV may reduce dose to the opposite eye wall by more than a factor of 2 while maintaining tolerable proximal sclera doses when prescribing to depths of 9 mm or less. Current commercially-available brachytherapy sources can exhibit up to 15% relative dosimetric sensitivity to seed design at regions within the eye.
本研究探讨了源光子能量对填充有(103)钯、(125)碘或(131)铯源或能量范围为12 keV至100 keV的单能光子发射的16毫米COMS眼敷贴近距离治疗剂量分布的影响。针对三种常见近距离治疗籽源设计的所有排列组合,同样创建了剂量分布。在此范围内,平均能量≤22 keV的源在向9毫米或更浅深度处方时,可将对侧眼壁的剂量降低2倍以上,同时保持可耐受的近侧巩膜剂量。目前市售的近距离治疗源在眼内区域对籽源设计可表现出高达15%的相对剂量学敏感性。