Suppr超能文献

Fracture risk and initial fixation of a cementless glenoid implant: the effect of numbers and types of screws.

作者信息

Suarez Daniel R, Valstar Edward R, Rozing Piet M, van Keulen Fred

机构信息

Departamento de Ingeniería Industrial, Facultad de Ingeniería, Pontificia Universidad Javeriana, Bogotá, Colombia.

出版信息

Proc Inst Mech Eng H. 2013 Oct;227(10):1058-66. doi: 10.1177/0954411913491050. Epub 2013 Jun 26.

Abstract

The initial fixation of an anatomical cementless glenoid component, provided by different numbers and types of screws, and the risk of bone fracture were evaluated by estimating the bone-implant interface micromotions and the principal strains around the prosthesis. Four different fixation configurations using locking or compression screws were tested. Estimation of the micromotions at the bone-implant interface was performed both experimentally, using an in vitro model, and computationally, using a numerical model. Principal bone strains were estimated using the numerical model. Subject variability was included by modelling two different bone qualities (healthy and rheumatoid bone). For the fixation configurations that used two screws, experimental and modelling results found that the micromotions at the bone-implant interface did not change with screw type. However, screw type had a significant effect on fixation when only one screw was used; in this case, a locking screw resulted in less micromotion at the bone-implant interface compared with the compression screw. Bone strains were predicted by the numerical model, and strains were found to be independent of the screw type; however, the predicted strain levels calculated in rheumatoid bone were larger than the strain levels that may cause bone damage for most considered arm positions. Predicted bone strain in healthy bone did not reach this level. While proper initial component fixation that allows biological fixation can be achieved by using additional screws, the risk of bone failure around the screws must be considered, especially in cases of weak bone.

摘要

相似文献

1
Fracture risk and initial fixation of a cementless glenoid implant: the effect of numbers and types of screws.
Proc Inst Mech Eng H. 2013 Oct;227(10):1058-66. doi: 10.1177/0954411913491050. Epub 2013 Jun 26.
2
Bone remodelling around a cementless glenoid component.
Biomech Model Mechanobiol. 2012 Jul;11(6):903-13. doi: 10.1007/s10237-011-0360-9. Epub 2011 Nov 23.
3
An experimental glenoid rim strain analysis for an improved reverse anatomy shoulder implant fixation.
J Orthop Res. 2012 Jun;30(6):998-1003. doi: 10.1002/jor.22015. Epub 2011 Nov 16.
4
Prediction at long-term condyle screw fixation of temporomandibular joint implant: A numerical study.
J Craniomaxillofac Surg. 2015 May;43(4):469-74. doi: 10.1016/j.jcms.2015.02.013. Epub 2015 Feb 24.
5
The effect of screw position on the initial fixation of a reverse total shoulder prosthesis in a glenoid with a cavitary bone defect.
J Shoulder Elbow Surg. 2008 May-Jun;17(3):479-86. doi: 10.1016/j.jse.2007.09.002. Epub 2008 Feb 20.
6
Factors contributing to glenoid baseplate micromotion in reverse shoulder arthroplasty: a biomechanical study.
J Shoulder Elbow Surg. 2019 Apr;28(4):648-653. doi: 10.1016/j.jse.2018.09.012. Epub 2018 Nov 30.
7
Does glenoid baseplate geometry affect its fixation in reverse shoulder arthroplasty?
J Shoulder Elbow Surg. 2012 Jul;21(7):917-24. doi: 10.1016/j.jse.2011.04.017. Epub 2011 Jul 20.
10
Locked plate fixation of osteoporotic humeral shaft fractures: are two locking screws per segment enough?
J Orthop Trauma. 2010 Apr;24(4):207-11. doi: 10.1097/BOT.0b013e3181bdd1da.

引用本文的文献

1
Biomechanical behaviours of the bone-implant interface: a review.
J R Soc Interface. 2019 Jul 26;16(156):20190259. doi: 10.1098/rsif.2019.0259. Epub 2019 Jul 31.
2
Mechanical and Biological Advantages of a Tri-Oval Implant Design.
J Clin Med. 2019 Mar 28;8(4):427. doi: 10.3390/jcm8040427.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验