Meyer Olivier, Greig Gerard, Holzgrefe Henry H
Institute of Clinical Pharmacology, F. Hoffmann-La Roche, Strasbourg, France.
J Electrocardiol. 2013 Nov-Dec;46(6):717.e1-9. doi: 10.1016/j.jelectrocard.2013.06.002. Epub 2013 Jul 11.
Technological advances in machine-read QT measurement now enable detailed and precise cardiac repolarization assessments. This study assessed the applicability of three state-of-art ECG measurement applications to provide reliable continuous analyses from data obtained in a positive thorough QT study previously characterized with sparse semi-automated measurements performed by an ECG core laboratory.
Continuous RR, QT, QTc measurements, and individual QT/RR relationships and their associated intra- and inter-subject variability were derived in parallel with BioQT, Ponemah PRO, and WinAtrec analysis software.
Despite slight vendor-specific differences in measurement variability and QTc, all machine-read methods demonstrated requisite assay sensitivity and yielded similar conclusions in accordance with SA analysis.
Three commercially available ECG analytical software applications reliably detected the drug-induced QT prolonging effects and replicated the SA core-laboratory conclusions, with greatly improved temporal resolution and reduced analytical costs. With broader experience, these data suggest that current SA methodologies could be effectively replaced by fully automated ECG analysis.