IEEE Trans Biomed Circuits Syst. 2011 Apr;5(2):138-46. doi: 10.1109/TBCAS.2011.2144592.
A wirelessly powered electro-acupuncture (EA) system with adaptive-pulsewidth (APW) monophase stimulation is presented for convenient invasive medicine. The proposed system removes cumbersome wires connected between EA nodes and an EA controller in order to realize both patients' convenience and remedial values simultaneously. An ultra-low-power stimulator integrated circuit (IC) that is integrated on the flexible-printed-circuit board (F-PCB) is attached to the tip of a needle electrode. Combined with a conductive yarn helical antenna wound around the needle electrode, the EA node receives wireless power from the EA controller using 433 MHz with the maximum loss of 6 dB. A zero-Vth nMOS rectifier harvests a supply voltage of 1.0 V from a -16-dBm incoming power signal with 32% efficiency. To deal with a body impedance variation (BIV) in the range of 100-200 kΩ , the proposed APW stimulator IC, fabricated in a 0.18-μm 1P6M complementary metal-oxide semiconductor CMOS process and occupying 1.56 mm(2), enables constant charge injection of 80-nC/stimulation. To ensure the patients' safety, the EA node (a pair of EAs) shares ground and clock wires to operate in alternate monophase (AMP) fashion for neutralizing the injected charge. The proposed wirelessly powered EA node was verified by applying it to a chunk of pork as a body model with the wireless power supplied from an RF signal generator (output power of 10 dBm and located 30 cm away).
提出了一种具有自适应脉冲宽度(APW)单相刺激的无线供电电针(EA)系统,用于方便的侵入性医学。所提出的系统去除了 EA 节点与 EA 控制器之间连接的繁琐电线,以同时实现患者的便利性和治疗价值。一个集成在柔性印刷电路板(F-PCB)上的超低功耗刺激器集成电路(IC)被附接到针尖电极上。结合缠绕在针尖电极周围的导电纱螺旋天线,EA 节点使用 433MHz 从 EA 控制器接收无线功率,最大损耗为 6dB。零 VthnMOS 整流器从 -16dBm 的输入功率信号中以 32%的效率获取 1.0V 的电源电压。为了应对 100-200kΩ 范围内的体阻抗变化(BIV),所提出的 APW 刺激器 IC 采用 0.18μm1P6M 互补金属氧化物半导体 CMOS 工艺制造,占用 1.56mm²,可实现 80-nC/刺激的恒定电荷注入。为了确保患者的安全,EA 节点(一对 EA)共享地和时钟线,以交替单相(AMP)方式运行,用于中和注入的电荷。所提出的无线供电 EA 节点已通过将其应用于作为体模型的猪肉块进行了验证,无线功率由射频信号发生器提供(输出功率为 10dBm,位于 30cm 之外)。