* Department of Experimental Psychology, University of Cambridge , UK.
Int J Audiol. 2013 Oct;52(10):678-86. doi: 10.3109/14992027.2013.809483. Epub 2013 Jul 17.
To compare the effectiveness of two types of treatment for unilateral hearing loss (UHL), bone-anchored hearing instruments (BAHI) and a dental device (SoundBite).
Either BAHI or SoundBite were worn for 30 days, and then the devices were swapped and the second device was worn for 30 days. Measures included unaided and aided sound-field thresholds, sound localization, and perception of speech in babble. The APHAB questionnaire was administered for each trial period.
Nine adult BAHI wearers with UHL.
Mid-frequency aided thresholds were lower for SoundBite than for BAHI. Both devices gave benefits for localization after 30 days, but there was no difference between devices. Speech perception was better for both devices than for unaided listening when the target speech came from the poorer hearing side or in front, and the interfering babble came from the better-hearing side. There was no consistent difference between devices. APHAB scores were better for SoundBite than for BAHI.
Speech perception and sound localization were similar for the two types of device, but the SoundBite led to lower aided thresholds and better APHAB scores than the BAHI.
比较两种单侧听力损失(UHL)治疗方法的疗效,骨锚式助听器(BAHI)和牙科设备(SoundBite)。
佩戴 BAHI 或 SoundBite 30 天,然后交换设备并佩戴第二个设备 30 天。测量指标包括未助听和助听声场阈值、声源定位和噪声下言语感知。在每个试验期都进行 APHAB 问卷评估。
9 名单侧听力损失的成年 BAHI 佩戴者。
SoundBite 的中频助听听阈比 BAHI 低。两种设备佩戴 30 天后对声源定位都有帮助,但设备之间无差异。当目标言语来自听力较差的一侧或前方,干扰噪声来自听力较好的一侧时,两种设备的言语感知均优于未助听,且都优于单独使用设备。设备之间无明显差异。SoundBite 的 APHAB 评分优于 BAHI。
两种设备的言语感知和声源定位效果相似,但 SoundBite 助听听阈比 BAHI 低,APHAB 评分也更高。