Zhang Weihua, Wang Zhiqiang, Liu Yina, Li Chunjuan, Xiao Xuefu, Luo Hailong, Chen Jun, Li Wei
China Institute of Atomic Energy, Beijing 102413, China.
Radiat Prot Dosimetry. 2014 Jan;158(2):246-50. doi: 10.1093/rpd/nct221. Epub 2013 Sep 13.
A portable tissue-equivalent proportional counter (TEPC)-based system was developed, which will be used as a reference standard in the neutron mixed field for radiation protection. In this paper, microdosimetry of the TEPC system was studied both in neutron ((252)Cf and (241)Am-Be sources) and gamma ((137)Cs and (60)Co sources) reference radiation fields. The measured neutron and gamma-dose equivalent rates were, respectively, compared with those of the reference ambient dose equivalent rate H(*)(10). And the measured microdosimetric spectra were compared with the Monte Carlo code FLUKA simulated results. All of the comparisons primarily agreed well.
开发了一种基于便携式组织等效正比计数器(TEPC)的系统,该系统将用作中子混合场辐射防护的参考标准。本文研究了TEPC系统在中子(²⁵²Cf和²⁴¹Am-Be源)和γ(¹³⁷Cs和⁶⁰Co源)参考辐射场中的微剂量学。将测得的中子和γ剂量当量率分别与参考环境剂量当量率H*(10)进行比较。并将测得的微剂量学谱与蒙特卡罗代码FLUKA模拟结果进行比较。所有比较结果基本吻合良好。