Fukuyama Yasuhiro, Suzuki Ryutaro, Takayama Shunsuke, Ueno Akinori
Annu Int Conf IEEE Eng Med Biol Soc. 2013;2013:555-8. doi: 10.1109/EMBC.2013.6609560.
An electrode configuration composed of multi-layered fabric electrodes were examined and compared with that composed of non-layered fabric electrodes in terms of movement artifact reduction when using the electrodes in capacitive electrocardiogram measurement from human buttocks in a vibrating environment. Experiments performed with six participants revealed that the multi-layered configuration composed of sensing electrode, driven shield and ground layers reduced amplitude of movement artifact significantly than the non-layered configuration.
在振动环境下从人体臀部进行电容式心电图测量时,研究并比较了由多层织物电极组成的电极配置与由非层状织物电极组成的电极配置在减少运动伪影方面的情况。对六名参与者进行的实验表明,由传感电极、驱动屏蔽层和接地层组成的多层配置比非层状配置能显著降低运动伪影的幅度。