Perperidis Antonios, Murray Ian, Brash Harry, McTrusty Alice, Cameron Lorraine, Fleck Brian, Minns Robert
Annu Int Conf IEEE Eng Med Biol Soc. 2013;2013:1636-9. doi: 10.1109/EMBC.2013.6609830.
The accurate assessment of visual field function can provide valuable information on a range of visual disorders. Saccadic Vector Optokinetic Perimetry (SVOP) is a novel instrument for measuring supra-threshold visual fields in young children who are otherwise unable to perform Automated Static Perimetry (ASP). However, limitations in Liquid Crystal Display (LCD) technology restrict the ability of SVOP to determine threshold values at various points in the visual field, often required in detailed perimetry examinations. This paper introduces a purpose-specific LCD luminance non-uniformity compensation approach to address this limitation. Thorough quantitative evaluation identifies the effectiveness of the proposed approach in (i) compensating for luminance non-uniformities across an LCD, and (ii) enabling SVOP to perform accurate and precise threshold visual field tests. The findings demonstrate that SVOP provides a promising alternative to the current threshold ASP standard (Humphrey Field Analyser).
对视野功能的准确评估能够为一系列视觉障碍提供有价值的信息。扫视矢量视动视野计(SVOP)是一种用于测量那些无法进行自动静态视野计(ASP)检测的幼儿超阈值视野的新型仪器。然而,液晶显示器(LCD)技术的局限性限制了SVOP在视野中各个点确定阈值的能力,而这在详细的视野检查中通常是需要的。本文介绍了一种针对特定目的的LCD亮度不均匀性补偿方法来解决这一局限性。全面的定量评估确定了所提出方法在以下两方面的有效性:(i)补偿LCD上的亮度不均匀性,以及(ii)使SVOP能够进行准确且精确的阈值视野测试。研究结果表明,SVOP为当前的阈值ASP标准(Humphrey视野分析仪)提供了一种有前景的替代方案。