We studied the feasibility of the precise size measurement of extraocular muscles using fast magnetic resonance imaging sequences and gaze fixation targets.
We recruited 20 healthy volunteers and optimized fat-suppressed fast T1-weighted gradient-recalled echo (FS T1-GRE) and single-shot fast spin-echo (SSFSE) imaging for evaluating extraocular muscles. With each eye at 40° abduction and adduction, we scanned the short-axis plane at the maximum diameter of the bilateral lateral and medial rectus muscles and measured the maximal cross-sectional area of the muscles during extension and contraction. We evaluated interobserver reproducibility and variability between the size measurements using the two MR sequences and the measurement reproducibility using FS T1-GRE imaging.
The interobserver reproducibility in the muscle size measurements using FS T1-GRE and SSFSE imaging was excellent (r = 0.93-0.94) and moderate to good (r = 0.54-0.75), respectively. The interobserver variability was less with FS T1-GRE than SSFSE imaging (p < 0.01). The reproducibility of the size measurement using FS T1-GRE was good to excellent (r = 0.78-0.92).
FS T1-GRE imaging with the subject staring at each of multiple targets is useful for evaluating precisely the size of extraocular muscles.
我们研究了使用快速磁共振成像序列和注视固定目标精确测量眼外肌大小的可行性。
我们招募了 20 名健康志愿者,并优化了用于评估眼外肌的脂肪抑制快速 T1 加权梯度回波(FS T1-GRE)和单次激发快速自旋回波(SSFSE)成像。在每只眼外展和内收 40°的情况下,我们在双侧直肌和内直肌最大直径处扫描短轴平面,并在伸展和收缩时测量肌肉的最大横截面积。我们使用两种 MR 序列评估了观察者间肌肉大小测量的可重复性和变异性,并使用 FS T1-GRE 成像评估了测量的可重复性。
使用 FS T1-GRE 和 SSFSE 成像测量肌肉大小的观察者间可重复性均为极好(r=0.93-0.94)和中度至良好(r=0.54-0.75)。FS T1-GRE 成像的观察者间变异性小于 SSFSE 成像(p<0.01)。使用 FS T1-GRE 进行的大小测量的重复性为良好至极好(r=0.78-0.92)。
受试者注视多个目标中的每一个时,FS T1-GRE 成像对于精确评估眼外肌的大小非常有用。