JAMA Otolaryngol Head Neck Surg. 2013 Dec;139(12):1343-50. doi: 10.1001/jamaoto.2013.5338.
Efficient diagnosis and early treatment of obstructive sleep apnea may help prevent the development of related morbidity and mortality. Compared with polysomnography (PSG), ambulatory sleep study devices offer the possibility of an accurate diagnosis with convenience and low cost.
To assess the correlation between sleep indexes measured by a portable sleep-testing device (peripheral arterial tonometry [PAT]) and those measured by PSG.
We searched PubMed, MEDLINE, the Cochrane Trial Registry (through May 2013), and relevant article bibliographies.
Systematic review and meta-analysis of studies assessing correlation of sleep indexes between PAT devices and PSG in adults (aged >18 years). Included studies provided a bivariate correlation coefficient for sleep indexes, specifically the respiratory disturbance index (RDI), apnea-hypopnea index (AHI), and oxygen desaturation index (ODI).
Included studies were reviewed by 2 independent reviewers. Reported correlation values for the RDI, AHI, and ODI between a commercially available PAT device (WatchPAT) and PSG were systematically reviewed. A comprehensive meta-analysis software package was used for statistical analysis.
Assessment of the correlation between PAT and PSG as measured by AHI, RDI, and ODI.
Fourteen studies met inclusion criteria and had data suitable for pooling (909 patients). Of these, 13 studies had blinded study designs, with PAT and PSG conducted simultaneously in the home or the laboratory setting. One study contained 2 trial phases for the same patient group (n = 29), one laboratory based and the other home based, which were analyzed separately. One study contained 2 different study groups based on age. Overall, correlation of the RDI and AHI was high (r = 0.889 [95% CI, 0.862-0.911]; P < .001). Studies comparing the RDI between PAT and PSG had a combined correlation of 0.879 (95% CI, 0.849-0.904; P < .001); those comparing the AHI, 0.893 (0.857-0.920; P < .001); and those comparing the ODI, 0.942 (0.894-0.969; P < .001). Analysis of publication bias revealed a nonsignificant Egger regression intercept.
Respiratory indexes calculated using PAT-based portable devices positively correlated with those calculated from the scoring of PSG. Strengthened by the blinded design of most of the included studies, this technology represents a viable alternative to PSG for confirmation of clinically suspected sleep apnea.
阻塞性睡眠呼吸暂停的有效诊断和早期治疗可能有助于预防相关发病率和死亡率的发生。与多导睡眠图(PSG)相比,便携式睡眠研究设备具有准确诊断的可能性,并且具有方便和低成本的优点。
评估便携式睡眠测试设备(外周动脉张力测定法 [PAT])测量的睡眠指标与 PSG 测量的睡眠指标之间的相关性。
我们在 PubMed、MEDLINE、Cochrane 试验注册库(截至 2013 年 5 月)和相关文章的参考文献中进行了搜索。
系统评价和荟萃分析评估 PAT 设备与成人 PSG 之间睡眠指标相关性的研究(年龄>18 岁)。纳入的研究提供了睡眠指标的双变量相关系数,特别是呼吸紊乱指数(RDI)、呼吸暂停低通气指数(AHI)和氧减饱和指数(ODI)。
由 2 名独立评审员对纳入的研究进行了审查。系统地审查了在商业上可用的 PAT 设备(WatchPAT)和 PSG 之间 RDI、AHI 和 ODI 的报告相关值。使用综合荟萃分析软件包进行统计分析。
评估 PAT 和 PSG 之间的相关性,以 AHI、RDI 和 ODI 进行测量。
14 项研究符合纳入标准,并具有适合汇总的数据(909 名患者)。其中,13 项研究采用了盲法设计,在家中或实验室同时进行 PAT 和 PSG。一项研究包含同一患者组的两个试验阶段(n=29),一个基于实验室,另一个基于家庭,分别进行分析。一项研究包含基于年龄的两组不同的研究群体。总体而言,RDI 和 AHI 的相关性很高(r=0.889 [95%CI,0.862-0.911];P<0.001)。比较 PAT 和 PSG 之间 RDI 的研究合并相关性为 0.879(95%CI,0.849-0.904;P<0.001);比较 AHI 的研究,0.893(0.857-0.920;P<0.001);比较 ODI 的研究,0.942(0.894-0.969;P<0.001)。对发表偏倚的分析显示,Egger 回归截距无统计学意义。
使用基于 PAT 的便携式设备计算的呼吸指标与从 PSG 评分计算的呼吸指标呈正相关。由于纳入的大多数研究都采用了盲法设计,因此该技术代表了一种替代 PSG 用于确认临床疑似睡眠呼吸暂停的可行方法。