利用离子阱串联质谱技术对粗血浆提取物进行定量分析。

Quantitative analysis of crude plasma extracts by ion trap tandem mass spectrometry.

机构信息

Miami Valley Laboratories, The Procter and Gamble Co., P.O. Box 398707, 45239-8707, Cincinnati, OH.

出版信息

J Am Soc Mass Spectrom. 1992 May;3(4):372-7. doi: 10.1016/1044-0305(92)87065-7.

Abstract

The ion trap mass spectrometer is a tandem-in-time instrument that has promise as an extremely sensitive device for practical tandem mass spectrometry assays. An approach for the quantitative analysis of unknown drug levels in crude extracts, using combined capillary gas chromatography and the ion trap mass spectrometer in the tandem mode, is described. One-gram plasma samples were spiked with an anti-inflammatory drug at levels of 1-100 ng, and with 50 ng of a chemical analog internal standard. Crude extracts of the plasma samples are analyzed by using scan functions that utilize combined radiofrequency (rf) and de voltages. The need for combined rf- and de-voltage sequences for analysis of such extracts is demonstrated by comparison to attempted analyses using only rf voltages. Limitations of the method are: (1) the need for accurate calibration of ionization times to obtain linear calibration lines, and (2) the lack of automatic gain control for scans using combined rf and dc voltages to control and optimize parent ion populations and to allow a simpler analysis of "unknowns. "

摘要

离子阱质谱仪是一种串联式时间仪器,有望成为一种非常灵敏的实用串联质谱分析装置。本文描述了一种在串联模式下,结合毛细管气相色谱和离子阱质谱仪,对粗提物中未知药物水平进行定量分析的方法。将 1 至 100ng 的抗炎药物和 50ng 的化学模拟物内标加入到 1 克血浆样品中,对粗提物进行分析,采用同时利用射频(rf)和直流(dc)电压的扫描功能。通过与仅使用射频电压的尝试分析进行比较,证明了这种提取方法需要同时使用射频和直流电压序列。该方法的局限性在于:(1)需要准确校准离子化时间以获得线性校准曲线;(2)在使用射频和直流电压组合进行扫描时缺乏自动增益控制,无法控制和优化母离子群体,也无法简化对“未知物”的分析。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验