We describe a previously unreported pitfall, spread of the stimulus at the elbow to the radial nerve, in an antidromic sensory nerve conduction study of the lateral antebrachial cutaneous (LAC) nerve.
Subjects consisted of 80 healthy volunteers, and both sides were examined for each subject. Besides routine recording of the LAC nerve, sensory nerve action potentials (SNAPs) of the radial nerve were recorded distally.
The spread phenomenon occurred in 73 of 160 arms (46%), and the SNAP amplitude increased due to contamination of the radial SNAP up to 6.7 times the genuine LAC SNAP. In 10 arms (6%), the spread started before the LAC SNAP was saturated, and the genuine LAC SNAP was unknown due to an anatomical variation in at least 1 arm.
Without monitoring distal radial SNAPs, the spread phenomenon will remain unrecognized. This pitfall undermines the reliability of the test.
我们描述了一种以前未报道的陷阱,即在逆行感觉神经传导研究中,肘部刺激向桡神经扩散,这种现象发生在外侧前臂皮神经(LAC)中。
研究对象包括 80 名健康志愿者,每位受试者的两侧均进行了检查。除了常规记录 LAC 神经外,还在远端记录桡神经感觉神经动作电位(SNAP)。
在 160 只手臂中,73 只(46%)出现了扩散现象,由于桡神经 SNAP 的干扰,SNAP 幅度增加了 6.7 倍。在 10 只手臂中(6%),扩散现象先于 LAC SNAP 饱和开始,由于至少 1 只手臂的解剖变异,导致真正的 LAC SNAP 无法识别。
如果不监测远端桡神经 SNAP,这种扩散现象将无法被识别。这个陷阱降低了测试的可靠性。