Gambarini Gianlucca, Gergi Richard, Grande Nicola Maria, Osta Nada, Plotino Gianluca, Testarelli Luca
Department of Endodontics, La Sapienze University, Rome, Italy Department of Endodontics, Saint Joseph University, Beirut, Lebanon.
Aust Endod J. 2013 Dec;39(3):151-4. doi: 10.1111/j.1747-4477.2012.00353.x. Epub 2012 May 2.
The aim of this study was to investigate whether cyclic fatigue resistance is increased for nickel titanium instruments manufactured with improved heating processes in clockwise or counterclockwise continuous rotation. The instruments compared were produced either using the R-phase heat treatment (K3XF; SybronEndo, Orange, CA, USA) or the M-wire alloy (ProFile Vortex; DENTSPLY Tulsa Dental Specialties, Tulsa, OK, USA). Tests were performed with a specific cyclic fatigue device that evaluated cycles to failure of rotary instruments in curved artificial canals. Results indicated no significant difference in resistance to cyclic fatigue when rotary nickel titanium instruments are used in clockwise or counterclockwise continuous rotation. In both directions of rotation, size 04-25 K3XF showed a significant increase (P < 0.05) in the mean number of cycles to failure when compared with size 04-25 ProFile Vortex.
本研究的目的是调查采用改进加热工艺制造的镍钛器械在顺时针或逆时针连续旋转时,其抗循环疲劳性能是否会提高。所比较的器械要么采用R相热处理(K3XF;美国加利福尼亚州奥兰治市SybronEndo公司)制造,要么采用M丝合金(ProFile Vortex;美国俄克拉何马州塔尔萨市DENTSPLY Tulsa Dental Specialties公司)制造。使用特定的循环疲劳装置进行测试,该装置评估旋转器械在弯曲人工根管中直至失效的循环次数。结果表明,当旋转镍钛器械顺时针或逆时针连续旋转时,其抗循环疲劳性能没有显著差异。在两个旋转方向上,与04 - 25号ProFile Vortex相比,04 - 25号K3XF的平均失效循环次数显著增加(P < 0.05)。