In vivo experimental porcine study.
To investigate if discography induced pressure increase in adjacent degenerate discs.
Despite refinements in the past 2 decades, the validity of discography is debated. Discography in healthy pigs has shown that the pressure increase during disc injection transmits to adjacent discs, a potential source for false positive responses.
Degeneration in 1 lumbar disc was induced in 10 pigs by drilling a hole through the endplate. Intradiscal pressure was recorded using a 0.36-mm fiber-optic pressure transducer inserted into nucleus pulposus through a 22-gauge needle. The pressure was measured simultaneously in 2 adjacent discs during slow (0.03 mL/s) automated contrast injection into 1 of the discs up to 8 bar (116 psi). Ten adjacent discs were prefilled with contrast from previous discogram. A pressure increase 2 psi or more above baseline was defined as increased pressure in adjacent discs. Pressure was recorded until 15 minutes after injection.
A total of 28 discograms were successfully performed. A pressure increase during injection was detected in 57% (16) of the 28 adjacent discs with mean 3.2 psi (1.7-8.2; standard deviation, 1.8), corresponding to a mean increase above baseline of 11%. Of those 16 adjacent discs, 4 were nondegenerate and 12 degenerate, of which 7 were prefilled. Fifteen minutes after injection, 89% of adjacent discs displayed increase in pressure of mean 14% above baseline.
Discography induced pressure increase in adjacent discs in a degenerate disc model, something not reported earlier. If present, also in human spine pressure transmission, may be a potential cause for false positive discography responses.
体内实验性猪研究。
研究椎间盘造影是否会导致相邻退变椎间盘内压力升高。
尽管在过去的 20 年中进行了改进,但椎间盘造影的有效性仍存在争议。在健康猪的椎间盘造影中已经表明,椎间盘注射过程中的压力增加会传递到相邻的椎间盘,这可能是假阳性反应的潜在原因。
通过在终板上钻孔,在 10 头猪的 1 个腰椎间盘内诱导退变。通过插入核髓质的 22 号针中的 0.36 毫米光纤压力传感器记录椎间盘内压力。在 1 个椎间盘缓慢(0.03 毫升/秒)自动对比注射至 8 巴(116 磅/平方英寸)的过程中,同时测量 2 个相邻椎间盘的压力。10 个相邻椎间盘在之前的椎间盘造影中已用对比剂预填充。将压力增加 2 磅/平方英寸或以上定义为相邻椎间盘的压力增加。记录压力直至注射后 15 分钟。
总共成功进行了 28 次椎间盘造影。在 28 个相邻椎间盘中有 57%(16 个)在注射过程中检测到压力增加,平均为 3.2 磅/平方英寸(1.7-8.2;标准差,1.8),相当于基线以上的平均增加 11%。在这 16 个相邻椎间盘中有 4 个是非退变的,12 个是退变的,其中 7 个是预填充的。注射后 15 分钟,89%的相邻椎间盘的压力增加,平均比基线增加 14%。
在退变椎间盘模型中,椎间盘造影会导致相邻椎间盘内压力升高,这在以前的研究中没有报道过。如果在人类脊柱中也存在压力传递,可能是椎间盘造影假阳性反应的潜在原因。