基于 EBT2 胶片的伽玛刀扩展系统分次治疗计划验证。

Verification of Gamma Knife extend system based fractionated treatment planning using EBT2 film.

机构信息

Gamma Knife Unit, Department of Neurosurgery, Neurosciences Centre, All India Institute of Medical Sciences, Ansari Nagar, New Delhi 110029, India.

出版信息

Med Phys. 2013 Dec;40(12):122104. doi: 10.1118/1.4832138.

Abstract

PURPOSE

This paper presents EBT2 film verification of fractionated treatment planning with the Gamma Knife (GK) extend system, a relocatable frame system for multiple-fraction or serial multiple-session radiosurgery.

METHODS

A human head shaped phantom simulated the verification process for fractionated Gamma Knife treatment. Phantom preparation for Extend Frame based treatment planning involved creating a dental impression, fitting the phantom to the frame system, and acquiring a stereotactic computed tomography (CT) scan. A CT scan (Siemens, Emotion 6) of the phantom was obtained with following parameters: Tube voltage--110 kV, tube current--280 mA, pixel size--0.5 × 0.5 and 1 mm slice thickness. A treatment plan with two 8 mm collimator shots and three sectors blocking in each shot was made. Dose prescription of 4 Gy at 100% was delivered for the first fraction out of the two fractions planned. Gafchromic EBT2 film (ISP Wayne, NJ) was used as 2D verification dosimeter in this process. Films were cut and placed inside the film insert of the phantom for treatment dose delivery. Meanwhile a set of films from the same batch were exposed from 0 to 12 Gy doses for calibration purposes. An EPSON (Expression 10000 XL) scanner was used for scanning the exposed films in transparency mode. Scanned films were analyzed with inhouse written MATLAB codes.

RESULTS

Gamma index analysis of film measurement in comparison with TPS calculated dose resulted in high pass rates >90% for tolerance criteria of 1%∕1 mm. The isodose overlay and linear dose profiles of film measured and computed dose distribution on sagittal and coronal plane were in close agreement.

CONCLUSIONS

Through this study, the authors propose treatment verification QA method for Extend frame based fractionated Gamma Knife radiosurgery using EBT2 film.

摘要

目的

本文介绍了 EBT2 胶片在伽玛刀(GK)扩展系统分割治疗计划中的验证,该系统是一种用于多次分割或序贯多次立体定向放射外科的可移动框架系统。

方法

一个人头形状的体模模拟了分割伽玛刀治疗的验证过程。基于扩展框架的治疗计划的体模准备涉及创建牙印、将体模适配到框架系统以及获取立体定向计算机断层扫描(CT)扫描。使用以下参数对体模进行 CT 扫描(西门子,Emotion 6):管电压-110kV,管电流-280mA,像素大小-0.5×0.5 和 1mm 切片厚度。制作了一个 8mm 准直器射击和每个射击中的三个扇区阻挡的治疗计划。在计划的两个部分中的第一个部分中,以 4Gy 的剂量处方进行了 100%的剂量传递。在这个过程中,使用 Gafchromic EBT2 胶片(ISP 韦恩,NJ)作为 2D 验证剂量计。胶片切割并放置在体模的胶片插入物内,以进行治疗剂量传递。同时,从同一批胶片中取出一组胶片,用于 0 至 12Gy 剂量的校准。使用 EPSON(Expression 10000XL)扫描仪以透明模式扫描曝光胶片。使用内部编写的 MATLAB 代码分析扫描胶片。

结果

与 TPS 计算剂量相比,胶片测量的伽马指数分析导致高通过率>90%,容限标准为 1%∕1mm。胶片测量和计算剂量分布的等剂量线叠加和线性剂量曲线在矢状面和冠状面非常吻合。

结论

通过这项研究,作者提出了使用 EBT2 胶片对基于扩展框架的分割伽玛刀立体定向放射外科进行治疗验证 QA 的方法。