Lopez Julio, Vilallonga Ramon, Targarona Eduardo M, Balague Carmen, Enriquez Lenin, Rivera Ramon, Balibrea Jose M, Perez-Ochoa Francisco, Rodriguez Karime, Baeza Miguel, Reyes Arturo
Department of Surgery, HGZ 35, Mexican Institute for Social Security , Juarez , Mexico.
Minim Invasive Ther Allied Technol. 2014 Jun;23(3):144-51. doi: 10.3109/13645706.2013.875924. Epub 2013 Dec 20.
LigaSure™ was developed as an alternative to suture ligatures, hemoclips and staplers for ligating vessels and tissue bundles. The aim of the present study was to determine whether LigaSure can be used as a welding instrument in the performance of laparoscopic sleeve gastrectomy.
Gastric specimens were assigned into four groups. Group 1 - specimens remained with the staple line intact. Group 2 - the staple line was oversewn. Group 3 - the staple line was resected with LigaSure. Group 4 - staple line was resected with LigaSure and the seal was oversewn. In all specimens the pressure tolerance was assessed using a portable sensor.
In group 1 the leak pressure was 34.7 ± 11.7 whereas in group 2 specimens the pressure increased three-fold (101.9 ± 21.4). The LigaSure seal alone (group 3) achieved a mean pressure of 13.7 mmHg. However, in group 4 there was an exponential increase on their burst strength up to 142 mmHg (p = 0.0005).
According to our results, LigaSure could be used to perform laparoscopic sleeve gastrectomy with reduction of staple-line bleeding and, when reinforced with a running suture, it achieves a strength that approaches that of staples plus oversewing.
LigaSure™是作为缝合结扎、血管夹和吻合器的替代产品而开发的,用于结扎血管和组织束。本研究的目的是确定LigaSure能否用作腹腔镜袖状胃切除术的焊接器械。
胃标本分为四组。第1组——标本的吻合线保持完整。第2组——吻合线进行了连续缝合。第3组——用LigaSure切除吻合线。第4组——用LigaSure切除吻合线并对缝合处进行连续缝合。在所有标本中,使用便携式传感器评估耐压性。
第1组的漏压为34.7±11.7,而第2组标本的压力增加了两倍(101.9±21.4)。仅使用LigaSure进行缝合(第3组)的平均压力为13.7 mmHg。然而,第4组的破裂强度呈指数级增加,高达142 mmHg(p = 0.0005)。
根据我们的结果,LigaSure可用于进行腹腔镜袖状胃切除术,减少吻合线出血,并且当用连续缝合加强时,其强度接近吻合器加连续缝合的强度。