Int J Oral Maxillofac Implants. 2014 Jan-Feb;29(1):71-8. doi: 10.11607/jomi.3304.
The aim of this study was to assess and compare the visibility, diameter, and course of the mandibular incisive canal (MIC) using cone beam computed tomography (CBCT) and panoramic radiography.
CBCT images and panoramic radiographs from 243 patients were used in this study. Standard exposure and patient positioning protocols were used for all the patients. Both types of images were assessed by two dentomaxillofacial radiologists. The diameter and the endpoint level of the MIC were measured using the CBCT images. Statistical analysis was performed using t tests in statistical software.
Of the 486 hemimandibles examined, the MIC was visible in 249 (51.2%) radiographs and 459 (94.4%) CBCT images. The mean diameters of the MICs were 1.91±0.45 mm on the right side and 1.94±0.41 mm on the left side. The MICs on both the right and left sides of the mandible showed statistically significant differences in diameter in male versus female patients. The visibility of the MIC on the panoramic radiographs according to the increase in the diameter was not statistically significant for both sides. Twenty MICs reached to the midline of the mandible, and the majority of the MICs (n=114) terminated between the canine and the first premolar.
The visibility of the MIC in CBCT is much better than that observed in conventional panoramic radiography. Even some large MICs could not be observed in panoramic radiographs. Detection of the MIC using CBCT may be crucial for surgical procedures involving the interforaminal region.
本研究旨在评估和比较锥形束 CT(CBCT)和全景片对下颌切牙管(MIC)的可视性、直径和走行。
本研究使用了 243 例患者的 CBCT 图像和全景片。所有患者均采用标准曝光和患者定位方案。两位牙颌面放射科医生评估了这两种类型的图像。使用 CBCT 图像测量 MIC 的直径和终点水平。统计分析采用统计软件中的 t 检验。
在检查的 486 个半下颌中,249 个(51.2%)的全景片和 459 个(94.4%)的 CBCT 图像可见 MIC。右侧 MIC 的平均直径为 1.91±0.45mm,左侧 MIC 的平均直径为 1.94±0.41mm。男性和女性患者下颌左右两侧 MIC 的直径均存在统计学差异。两侧 MIC 在全景片上的可视性与直径的增加无统计学意义。20 个 MIC 延伸至下颌中线,大多数 MIC(n=114)终止于犬齿和第一前磨牙之间。
CBCT 中 MIC 的可视性明显优于常规全景片。即使一些较大的 MIC 也无法在全景片中观察到。使用 CBCT 检测 MIC 可能对涉及眶下区域的手术程序至关重要。