The suitability of a recently introduced inexpensive, compact mass spectrometer detector is evaluated for supporting pharmaceutical chemistry investigations. While high performance/high cost MS detectors dominate the marketplace, there is growing recognition of the need for a small, inexpensive MS detector with reduced capabilities for supporting synthetic chemistry investigations, where reduced sensitivity and unit mass resolution are often suitable for solving routine problems. In this study, the fundamental performance characteristics of the recently introduced Advion compact mass spectrometer were evaluated, investigating the use of the instrument for routine product and impurity identification, reaction monitoring, evaluation of potential genotoxic impurities and study of high molecular weight biomolecules. In general, the results of the evaluation show this compact and inexpensive mass spectrometer to be well suited for providing reliable support for pharmaceutical chemistry investigations, with sub-nanogram limit of detection and impurity identification below 0.1% being possible in some instance.
对最近推出的一种价格低廉、体积紧凑的质谱仪检测器在支持药物化学研究方面的适用性进行了评估。虽然高性能/高成本的质谱检测器在市场上占据主导地位,但人们越来越认识到需要一种功能有所降低的小型、廉价质谱检测器来支持合成化学研究,在合成化学研究中,较低的灵敏度和单位质量分辨率通常适合解决常规问题。在本研究中,对最近推出的Advion紧凑型质谱仪的基本性能特征进行了评估,考察了该仪器在常规产品和杂质鉴定、反应监测、潜在遗传毒性杂质评估以及高分子量生物分子研究中的应用。总体而言,评估结果表明,这种紧凑且廉价的质谱仪非常适合为药物化学研究提供可靠支持,在某些情况下,检测限可达亚纳克,杂质鉴定水平可低于0.1%。