Pavo Noemi, Syeda Bonni, Bernhart Andreas, Szentirmai Eszter, Hemetsberger Rayyan, Samaha Eslam, Plass Christian, Zlabinger Katrin, Pavo Imre J, Petrasi Zsolt, Petnehazy Örs, Hoerstrup Simon P, Maurer Gerald, Gyöngyösi Mariann
Department of Cardiology, Medical University of Vienna, Vienna, Austria.
EuroIntervention. 2015 Aug;11(4):433-41. doi: 10.4244/EIJV11I4A85.
To evaluate the short-, mid- and long-term safety, efficacy and vascular physiology of Axetis silicon dioxide (SiO2, abrading the micropores) inert-coated stent implantation in a randomised preclinical setting.
Coronary arteries of domestic pigs were randomised to receive either Axetis or BMS (same design) stents with one-, three- and six-month follow-up (FUP), controlled by coronary angiography, optical coherence tomography (OCT), intravascular ultrasound (IVUS) and histology (n=32). The time-dependent vasomotor reaction of coronary arteries to stenting was measured using modified myography (n=12). Complete endothelialisation of the Axetis stent was confirmed by OCT, IVUS and histology at one-month FUP. Histopathology revealed continuous healing of the vessel wall with a gradual reduction of inflammation and fibrin score during the six-month FUP in both stent types. Significantly smaller neointimal area and %area stenosis were measured in Axetis stents compared with BMS at each FUP time point. Vascular reactivity measurements showed significantly better endothelium-dependent vasodilation of stented arteries with Axetis implantation.
Implantation of the Axetis SiO2-coated stent resulted in a significantly better safety, efficacy and vessel physiology profile compared with BMS of the same design with a continuous decrease in vessel inflammation during the six-month FUP.
在随机临床前研究中评估Axetis二氧化硅(SiO2,磨蚀微孔)惰性涂层支架植入的短期、中期和长期安全性、有效性及血管生理学情况。
将家猪的冠状动脉随机分组,分别植入Axetis或BMS(相同设计)支架,并进行1个月、3个月和6个月的随访(FUP),通过冠状动脉造影、光学相干断层扫描(OCT)、血管内超声(IVUS)和组织学检查进行对照(n = 32)。使用改良肌动描记法测量冠状动脉对支架植入的时间依赖性血管舒缩反应(n = 12)。在1个月随访时,通过OCT、IVUS和组织学检查证实Axetis支架完全内皮化。组织病理学显示,在6个月随访期间,两种支架类型的血管壁均持续愈合,炎症和纤维蛋白评分逐渐降低。在每个随访时间点,Axetis支架的新生内膜面积和狭窄面积百分比均显著小于BMS。血管反应性测量显示,植入Axetis后,支架置入动脉的内皮依赖性血管舒张明显更好。
与相同设计的BMS相比,植入Axetis SiO2涂层支架的安全性、有效性和血管生理学情况显著更好,在6个月随访期间血管炎症持续减轻。