Entrapment of tined leads.

作者信息

Res J C, De Cock C C, Van Rossum A C, Schreuder J

机构信息

Department of Cardiology, Free University Hospital, Amsterdam, The Netherlands.

出版信息

Pacing Clin Electrophysiol. 1989 Oct;12(10):1583-5. doi: 10.1111/j.1540-8159.1989.tb01833.x.

Abstract

A rare complication is described, in three patients. The tined tip of a ventricular pacemaker electrode was entrapped in the chordae of the tricuspid valve and could not be removed by subtle manipulations in two patients. In one patient, the electrode was removed with partial rupture of the tricuspid valve. The two electrodes remained in the entrapped position and new electrodes were inserted in all three cases. No clinical sequelae were found during follow-up of at least 24 months.

摘要

本文描述了3例患者发生的一种罕见并发症。2例患者的心室起搏器电极尖端被三尖瓣腱索卡住,经精细操作仍无法取出。1例患者在取出电极时三尖瓣部分破裂。在所有3例中,两根电极均留在卡压位置,并植入了新的电极。在至少24个月的随访中未发现临床后遗症。