To evaluate the computed tomography in diagnosis of foreign bodies in the eye or close to the scleral wall, it was used with routine orbital radiography for comparison. The Simens Somatom DR3 CT Scanner with a 512 x 512 matrix was used in 50 random cases, and 45 cases were confirmed by surgery. The results indicated that the positive rate of CT in detecting intraocular foreign bodies was 98% (49/50), vs 90% (45/50) for radiography. The smallest metallic intraocular foreign body detected by CT was 0.4 x 0.25 mm and a cupric foreign body of 0.2 x 0.1 mm escaped detection. The error rate of CT in detecting single foreign body in the ocular wall was 1/30, while that of radiography was 10/30. Intraocular multiple foreign bodies could be demonstrated by multiple laminae of CT, and non-metallic foreign bodies could also be differentiated.
为评估计算机断层扫描在诊断眼内或靠近巩膜壁异物方面的应用,将其与常规眼眶X光摄影术进行比较。使用了具有512×512矩阵的西门西门子Somatom DR3 CT扫描仪对50例随机病例进行检查,其中45例经手术证实。结果表明,CT检测眼内异物的阳性率为98%(49/50),而X光摄影术的阳性率为90%(45/50)。CT检测到的最小金属眼内异物为0.4×0.25毫米,一个0.2×0.1毫米的铜质异物漏检。CT检测眼壁单个异物的错误率为1/30,而X光摄影术的错误率为10/30。多层CT可显示眼内多个异物,并且还能鉴别非金属异物。