The impact of the reference imaging modality, registration method and intraoperative flat-panel computed tomography on the accuracy of the ROSA® stereotactic robot.
作者信息
Lefranc Michel, Capel Cyril, Pruvot Anne Sophie, Fichten Anthony, Desenclos Christine, Toussaint Patrick, Le Gars Daniel, Peltier Johan
机构信息
Service de Neurochirurgie, CHU Amiens, Amiens, France.
出版信息
Stereotact Funct Neurosurg. 2014;92(4):242-50. doi: 10.1159/000362936. Epub 2014 Aug 20.
OBJECTIVE
To establish the impact of the imaging modality, registration method and use of intraoperative computed tomography (CT) scan on the accuracy of the ROSA® stereotactic robot.
METHODS
Using a dedicated phantom device, we measured the accuracy of the stereotactic robot for 20 targets as a function of the registration method (frameless, FL, or frame-based, FB) and the reference imaging modality (3T magnetic resonance imaging, MRI, CT scanner or flat-panel CT, fpCT). We performed a retrospective study of the accuracy of the first 26 FB and 31 FL robotized stereotactic surgeries performed in our department.
RESULTS
In a phantom study, the mean target accuracy was 1.59 mm for 3T MRI-guided FL surgery, 0.3 mm for fpCT-guided FL surgery and 0.3 mm for CT-guided FB surgery. In our retrospective series, the mean accuracy was 0.81 mm for FB stereotactic surgery, 1.22 mm for our 24 stereotactic surgery procedures with FL (surface recognition) registration and 0.7 mm for our 7 stereotactic surgery procedures with FL fiducial marker registration. Intraoperative fpCT fully corrected all the registration errors.
CONCLUSIONS
The ROSA stereotactic robot is highly accurate. Robotized FB stereotactic surgery is more accurate than robotized FL stereotactic surgery.