Suppr超能文献

Thoratec 搏动性心室辅助装置时机对腹主动脉波强度模式的影响。

Effects of Thoratec pulsatile ventricular assist device timing on the abdominal aortic wave intensity pattern.

机构信息

Institute for Dynamic Systems and Control, ETH Zurich, Zurich, Switzerland;

Department of Cardiology, University Hospital Bern, Bern, Switzerland;

出版信息

Am J Physiol Heart Circ Physiol. 2014 Oct 15;307(8):H1243-51. doi: 10.1152/ajpheart.00085.2014. Epub 2014 Aug 22.

Abstract

Arterial waves are seen as possible independent mediators of cardiovascular risks, and the wave intensity analysis (WIA) has therefore been proposed as a method for patient selection for ventricular assist device (VAD) implantation. Interpreting measured wave intensity (WI) is challenging, and complexity is increased by the implantation of a VAD. The waves generated by the VAD interact with the waves generated by the native heart, and this interaction varies with changing VAD settings. Eight sheep were implanted with a pulsatile VAD (PVAD) through ventriculoaortic cannulation. The start of PVAD ejection was synchronized to the native R wave and delayed between 0 and 90% of the cardiac cycle in 10% steps or phase shifts (PS). Pressure and velocity signals were registered, with the use of a combined Doppler and pressure wire positioned in the abdominal aorta, and used to calculate the WI. Depending on the PS, different wave interference phenomena occurred. Maximum unloading of the left ventricle (LV) coincided with constructive interference and maximum blood flow pulsatility, and maximum loading of the LV coincided with destructive interference and minimum blood flow pulsatility. We believe that noninvasive WIA could potentially be used clinically to assess the mechanical load of the LV and to monitor the peripheral hemodynamics such as blood flow pulsatility and risk of intestinal bleeding.

摘要

动脉波被视为心血管风险的可能独立介质,因此波强度分析(WIA)已被提议作为选择心室辅助装置(VAD)植入患者的方法。解释测量的波强度(WI)具有挑战性,并且由于植入 VAD 而增加了复杂性。VAD 产生的波与原生心脏产生的波相互作用,这种相互作用随 VAD 设置的变化而变化。通过心室 - 主动脉插管将 8 只绵羊植入搏动性 VAD(PVAD)。PVAD 射血的开始与原生 R 波同步,并在 0%至 90%的心脏周期内以 10%的步长或相位偏移(PS)延迟。使用位于腹主动脉中的组合多普勒和压力线记录压力和速度信号,并用于计算 WI。根据 PS 的不同,会发生不同的波干扰现象。左心室(LV)的最大卸载与建设性干扰和最大血流脉动性一致,而 LV 的最大加载与破坏性干扰和最小血流脉动性一致。我们认为,无创 WIA 可能具有临床应用潜力,可用于评估 LV 的机械负荷,并监测血流脉动性和肠道出血风险等外周血液动力学。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验