Jouyban-Gharamaleki Vahid, Jouyban-Gharamaleki Karim, Shayanfar Ali, Khoubnasabjafari Mehry, Jouyban Abolghasem
Department of Mechatronic Engineering, International Campus, University of Tabriz, Tabriz, Iran.
Tabriz Technical Training Center No. 2, Tabriz, Iran.
J Lab Autom. 2015 Feb;20(1):3-9. doi: 10.1177/2211068214554801. Epub 2014 Oct 20.
The aqueous solubility of a drug candidate is a vital physicochemical property that stops a drug candidate from proceeding further in the drug development processes. Classical solubility determination methods, which are commonly used in pharmaceutical laboratories, are expensive and time-consuming. In this work, an automated determination method is proposed that is based on a laser monitoring technique, and the validity of the measured solubilities is checked by comparing the measured solubilities of acetaminophen at various temperatures as proposed in various literatures. An additional set of acetaminophen solubilities in various concentrations of a surface active agent is measured at various temperatures, which has been reported for the first time, and it could be applied in the pharmaceutical industry, where solubilization of acetaminophen in aqueous solutions is required.
候选药物的水溶性是一种至关重要的物理化学性质,它会阻碍候选药物在药物研发过程中进一步推进。制药实验室常用的传统溶解度测定方法既昂贵又耗时。在这项工作中,提出了一种基于激光监测技术的自动测定方法,并通过比较不同文献中提出的对乙酰氨基酚在不同温度下的测量溶解度来检验所测溶解度的有效性。首次测量了对乙酰氨基酚在不同温度下于各种浓度表面活性剂中的另一组溶解度,这可应用于需要将对乙酰氨基酚溶解于水溶液的制药行业。