Institute of Academic Anaesthesia, University of Dundee, Dundee, UK.
Anaesthesia. 2015 May;70(5):537-42. doi: 10.1111/anae.12925. Epub 2014 Nov 5.
Visibility of the needle tip and shaft is important during ultrasound-guided regional anaesthesia in order to prevent nerve trauma. Tip and shaft visibility is reduced when needles are inserted in-plane at wide angles and out-of-plane at narrow angles to the ultrasound probe. Although textured needles are more reflective than smooth needles, we hypothesised that poor visibility of the tip and shaft still remained using the above angle-probe combinations. In a single-blind study, we compared the visibility of a textured Tuohy needle, a textured single-shot needle and a conventional smooth-surfaced Tuohy needle when inserted into the biceps and deltoid muscles of a soft embalmed cadaver. One hundred and forty-four needles were block-randomised to in-plane and out-of-plane insertions at 30°, 45°, 60° and 75° to the ultrasound beam. Two blinded raters assessed needle tip visibility on video recordings of the insertions using a binary scale (0 = not visible, 1 = visible) and shaft visibility using a 5-point Likert scale. The median (IQR [range]) proportions of visible needle tips were 83% (67-83 [50-100]%) for the textured Tuohy, 75% (67-83 [33-83]%) for the textured single-shot needle and 33% (33-46 [0-50]%) for the smooth-surfaced Tuohy (p = 0.0007). Median (IQR [range]) needle shaft visibility was rated as 4.0 (3.5-4.7 [3.0-4.9]) for the textured Tuohy, 4.0 (3.8-4.5 [2.7-4.9]) for the textured single-shot needle and 3.0 (2.4-3.3 [2.3-3.5]) for the smooth-surfaced Tuohy (p = 0.015). Nevertheless, visibility was reduced at wide angles in-plane and narrow angles out-of-plane both for needle tips (p = 0.004) and shafts (p = 0.005).
在超声引导下的区域麻醉中,为了防止神经损伤,针尖和针身的可视性很重要。当针尖和针身在与超声探头成大角度平面内插入和与探头成小角度平面外插入时,其可视性会降低。尽管纹理针比光滑针更具反射性,但我们假设在使用上述角度-探头组合时,针尖和针身的可视性仍然较差。在一项单盲研究中,我们比较了在软组织防腐尸体的肱二头肌和三角肌中插入时,纹理 Tuohy 针、纹理单次使用针和常规光滑表面 Tuohy 针的可视性。144 根针以 30°、45°、60°和 75°的角度随机分为平面内和平面外插入,并以二进制(0 = 不可见,1 = 可见)和 5 分李克特量表评估针尖端可视性。在插入过程的视频记录中,两名盲法评估者评估了针尖可视性,使用二进制(0 = 不可见,1 = 可见)和 5 分李克特量表评估针身可视性。可见针尖的中位数(IQR [范围])比例分别为纹理 Tuohy 83%(67-83 [50-100]%)、纹理单次使用针 75%(67-83 [33-83]%)和光滑表面 Tuohy 33%(33-46 [0-50]%)(p = 0.0007)。纹理 Tuohy 的针身可视性中位数(IQR [范围])为 4.0(3.5-4.7 [3.0-4.9]),纹理单次使用针为 4.0(3.8-4.5 [2.7-4.9]),光滑表面 Tuohy 为 3.0(2.4-3.3 [2.3-3.5])(p = 0.015)。然而,在大角度平面内和小角度平面外,无论是针尖(p = 0.004)还是针身(p = 0.005),可视性均降低。