Milletari Fausto, Belagiannis Vasileios, Navab Nassir, Fallavollita Pascal
Med Image Comput Comput Assist Interv. 2014;17(Pt 2):570-7. doi: 10.1007/978-3-319-10470-6_71.
We propose a method to perform automatic detection and tracking of electrophysiology (EP) catheters in C-arm fluoroscopy sequences. Our approach does not require any initialization, is completely automatic, and can concurrently track an arbitrary number of overlapping catheters. After a pre-processing step, we employ sparse coding to first detect candidate catheter tips, and subsequently detect and track the catheters. The proposed technique is validated on 2835 C-arm images, which include 39,690 manually selected ground-truth catheter electrodes. Results demonstrated sub-millimeter detection accuracy and real-time tracking performances.
我们提出了一种在C形臂荧光透视序列中对电生理(EP)导管进行自动检测和跟踪的方法。我们的方法不需要任何初始化,完全自动,并且可以同时跟踪任意数量的重叠导管。经过预处理步骤后,我们采用稀疏编码首先检测候选导管尖端,随后检测并跟踪导管。所提出的技术在2835张C形臂图像上得到验证,这些图像包括39690个手动选择的真实导管电极。结果表明检测精度达到亚毫米级且具有实时跟踪性能。