Unfavorable outcomes from trunnion fretting and corrosion damage have been reported in the literature, gross failures of tapers in primary total hip arthroplasties have been less frequently reported. We report on 5 patients, who presented with gross trunnion failures of modular metal-on-polyethylene or ceramic-on-polyethylene bearings from 5 implant manufacturers, all necessitating revision surgery. None of these patients had an antecedent history of trauma, and the majority presented with pain or instability. No common factor was identified that may be predictive of these type of failures. Since there were 5 different stem designs, this suggests that it may be a rare generic phenomenon occurring with multiple designs. Currently, further investigations are necessary, including retrieval analysis, to identify risk factors that may predispose to such failures.

摘要

文献中已报道了耳轴微动磨损和腐蚀损伤的不良后果,而初次全髋关节置换术中锥度的严重失效报道较少。我们报告了5例患者,他们均出现了来自5家植入物制造商的模块化金属对聚乙烯或陶瓷对聚乙烯轴承的耳轴严重失效,均需要进行翻修手术。这些患者均无前驱创伤史,大多数患者表现为疼痛或不稳定。未发现可预测此类失效的共同因素。由于有5种不同的柄设计,这表明这可能是一种在多种设计中出现的罕见的普遍现象。目前,需要进一步调查,包括取出分析,以确定可能易导致此类失效的风险因素。