Serrao Sebastiano, Lombardo Giuseppe, Desiderio Giovanni, Buratto Lucio, Schiano-Lomoriello Domenico, Pileri Marco, Lombardo Marco
Fondazione G.B. Bietti IRCCS, 00198 Rome, Italy.
CNR-IPCF Unit of Support of Cosenza, 87036 Rende, Italy ; Vision Engineering Italy Srl, 00198 Rome, Italy.
J Ophthalmol. 2014;2014:520713. doi: 10.1155/2014/520713. Epub 2014 Nov 20.
Purpose. To investigate the structure and irregularity of the capsulotomy cutting edges created by two femtosecond (FS) laser platforms in comparison with manual continuous circular capsulorhexis (CCC) using environmental scanning electron microscopy (eSEM). Methods. Ten anterior capsulotomies were obtained using two different FS laser cataract platforms (LenSx, n = 5, and Victus, n = 5). In addition, five manual CCC (n = 5) were obtained using a rhexis forceps. The specimens were imaged by eSEM (FEI Quanta 400, OR, USA). Objective metrics, which included the arithmetic mean deviation of the surface (Sa) and the root-mean-square deviation of the surface (Sq), were used to evaluate the irregularity of both the FS laser capsulotomies and the manual CCC cutting edges. Results. Several microirregularities were shown across the FS laser capsulotomy cutting edges. The edges of manually torn capsules were shown, by comparison of Sa and Sq values, to be smoother (P < 0.05) than the FS laser capsulotomy edges. Conclusions. Work is needed to understand whether the FS laser capsulotomy edge microirregularities, not seen in manual CCC, may act as focal points for the concentration of stress that would increase the risk of capsular tear during phacoemulsification as recently reported in the literature.
目的。使用环境扫描电子显微镜(eSEM),比较两种飞秒(FS)激光平台制作的晶状体囊切开术切口边缘的结构和不规则性与手动连续环形撕囊术(CCC)。方法。使用两种不同的FS激光白内障手术平台(LenSx,n = 5;Victus,n = 5)制作十个前囊切开术。此外,使用撕囊镊进行五次手动CCC(n = 5)。通过eSEM(FEI Quanta 400,美国俄勒冈州)对标本进行成像。使用包括表面算术平均偏差(Sa)和表面均方根偏差(Sq)在内的客观指标来评估FS激光囊切开术和手动CCC切口边缘的不规则性。结果。FS激光囊切开术切口边缘呈现出一些微不规则性。通过比较Sa和Sq值,发现手动撕开的囊膜边缘比FS激光囊切开术边缘更光滑(P < 0.05)。结论。正如最近文献报道的那样,需要开展研究来了解FS激光囊切开术边缘的微不规则性(手动CCC中未见到)是否可能作为应力集中的焦点,从而增加超声乳化过程中囊膜撕裂的风险。