Suppr超能文献

引入一种新型体外模型以表征经皮肾镜取石术系统的流体动力学效应。

Introducing a Novel In Vitro Model to Characterize Hydrodynamic Effects of Percutaneous Nephrolithotomy Systems.

作者信息

Mager Rene, Balzereit Christopher, Reiter Michael, Gust Kilian, Borgmann Hendrik, Hüsch Tanja, Nagele Udo, Haferkamp Axel, Schilling David

机构信息

1 Department of Urology and Paediatric Urology, University Hospital Frankfurt , Frankfurt, Germany .

2 ExperTeach GmbH , Dietzenbach, Germany .

出版信息

J Endourol. 2015 Aug;29(8):929-32. doi: 10.1089/end.2014.0854. Epub 2015 Feb 5.

Abstract

PURPOSE

To widen the understanding of the fluid mechanics during percutaneous nephrolithotomy (PCNL), a new in vitro model for empiric studies was developed. Our first aim was to characterize common PCNL systems with regard to their intrapyelocaliceal pressure and their irrigation flow volume as possible parameters of hydrodynamic invasiveness.

MATERIALS AND METHODS

The model is based on a watertight cylindrical cast. Low pressure miniaturized PCNL (miniPCNL) systems and conventional PCNL systems were tested. Intrapelvic pressure and the total of irrigation flow volume per minute were measured for each PCNL system at 10 predefined levels of irrigation pressure between 40 and 130 cm H2O.

RESULTS

Linear regression showed a linear relation between irrigation pressure and intrapelvic pressure. Compared with miniPCNL systems, conventional PCNL sheaths with closed Rutner sidearm generate significantly higher intrapelvic pressures. Opening the Rutner sidearm leads to a decrease in intrapelvic pressure to the level of the open mini PCNL systems. Polynomal regression revealed a square root relation between irrigation pressure and irrigation flow volume.

CONCLUSIONS

Our in vitro model is suitable for obtaining reliable and valid data of intrapelvic pressure and irrigation flow volume of commonly used PCNL systems. Conventional PCNL sheaths with opened Rutner sidearm provide intrapelvic pressure and irrigation volume similar to miniPCNL. Consequently, a relevant difference of the hydrodynamic impact between conventional and miniPCNL systems are not necessarily to be assumed. Regression analysis revealed a new starting point for further empiric research of fluid mechanics during PCNL.

摘要

目的

为了拓宽对经皮肾镜取石术(PCNL)期间流体力学的理解,开发了一种用于实证研究的新型体外模型。我们的首要目标是,就肾盂内压力和冲洗流量这两个可能作为流体动力侵袭性参数的指标,对常见的PCNL系统进行特性描述。

材料与方法

该模型基于一个防水的圆柱形铸模。对低压小型PCNL(miniPCNL)系统和传统PCNL系统进行了测试。在40至130 cm H2O之间的10个预定义冲洗压力水平下,测量每个PCNL系统的肾盂内压力和每分钟冲洗流量总和。

结果

线性回归显示冲洗压力与肾盂内压力之间存在线性关系。与miniPCNL系统相比,带有封闭Rutner侧臂的传统PCNL鞘会产生显著更高的肾盂内压力。打开Rutner侧臂会导致肾盂内压力降至开放的miniPCNL系统的水平。多项式回归揭示了冲洗压力与冲洗流量之间的平方根关系。

结论

我们的体外模型适用于获取常用PCNL系统的肾盂内压力和冲洗流量的可靠且有效的数据。带有开放Rutner侧臂的传统PCNL鞘提供的肾盂内压力和冲洗量与miniPCNL相似。因此,不一定认为传统PCNL系统和miniPCNL系统之间在流体动力影响方面存在显著差异。回归分析为PCNL期间流体力学的进一步实证研究揭示了一个新的起点。