Hashimoto Shogo, Noguchi Claudia Cecilia Yamamoto, Furutani Eiko
Annu Int Conf IEEE Eng Med Biol Soc. 2014;2014:4835-8. doi: 10.1109/EMBC.2014.6944706.
Treatment of type 1 diabetes consists of maintaining postprandial normoglycemia using the correct prandial insulin dose according to food intake. Nonetheless, it is hardly achieved in practice, which results in several diabetes-related complications. In this study we present a feedforward plus feedback blood glucose control system that considers the glycemic index of foods. It consists of a preprandial insulin bolus whose optimal bolus dose and timing are stated as a minimization problem, which is followed by a postprandial closed-loop control based on model predictive control. Simulation results show that, for a representative carbohydrate intake of 50 g, the present control system is able to maintain postprandial glycemia below 140 mg/dL while preventing postprandial hypoglycemia as well.