Zhu Zhenghan, Brooks James, Makevey Oleksandr, Kay Steven M, Besio Walter G
Annu Int Conf IEEE Eng Med Biol Soc. 2014;2014:5248-51. doi: 10.1109/EMBC.2014.6944809.
We have previously shown that tripolar concentric ring electrode (TCRE) Laplacian electroencephalography (tEEG) has significantly better signal-to-noise ratio, spatial resolution, and mutual information than disc electrode electroencephalography (EEG). This paper compares the EEG signals acquired simultaneously from the outer ring of the TCRE (oTCRE), shorting all three elements of the TCRE (sTCRE) and disc electrode (disc) concurrently from nearly the same location on the human scalp. We calculated the average correlation for the time series between each pair of signals and average coherence over the pass-band frequencies between all pairs of signals as well. All the correlations and coherences were above 0.99. The results suggest that the oTCRE can be used to record EEG concurrently with tEEG from the same sensor at the same location.
我们之前已经表明,三极同心环电极(TCRE)拉普拉斯脑电图(tEEG)在信噪比、空间分辨率和互信息方面比盘状电极脑电图(EEG)有显著更好的表现。本文比较了同时从TCRE的外环(oTCRE)、将TCRE的所有三个元件短路(sTCRE)以及盘状电极(盘)在人类头皮几乎相同位置同时采集的EEG信号。我们还计算了每对信号之间时间序列的平均相关性以及所有信号对之间通带频率上的平均相干性。所有相关性和相干性均高于0.99。结果表明,oTCRE可用于在同一位置从同一传感器与tEEG同时记录EEG。