Suppr超能文献

Ceraflex封堵器与Amplatzer封堵器用于继发孔型房间隔缺损封堵的多中心临床经验

Ceraflex versus Amplatzer occluder for secundum atrial septal defect closure. Multicenter clinical experience.

作者信息

Astarcioglu M A, Kalcik M, Sen T, Aykan A C, Gokdeniz T, Gursoy O M, Karakoyun S, Kulahcioglu S, Gunduz S, Kilit C, Oylumlu M, Amasyali B

机构信息

Department of Cardiology, Evliya Celebi Training and Research Hospital, Kutahya, Turkey,

出版信息

Herz. 2015 Apr;40 Suppl 2:146-50. doi: 10.1007/s00059-014-4192-0. Epub 2015 Feb 8.

Abstract

AIM

The Ceraflex atrial septal defect occluder is an alternative device to the Amplatzer septal occluder with some structural innovations including flexible connection, increased flexibility, and minimized amount of implant material. We evaluated the efficiency and safety of the Ceraflex septal occluder device in percutaneous closure of secundum atrial septal defects.

PATIENTS AND METHODS

This was a prospective, nonrandomized, multicenter study of patients undergoing transcatheter closure for an atrial septal defect with the Ceraflex and the Amplatzer septal occluder devices. A clinical evaluation and follow-up transthoracic echocardiography were performed at 1, 6, and 12 months.

RESULTS

Between 2010 and 2014, 125 patients underwent atrial septal defect closure with the Ceraflex septal occluder (n = 58) and the Amplatzer septal occluder (n = 67) under transesophageal echocardiography guidance. Patient characteristics, the stretched size of the defect, device size, and fluoroscopy time were similar between the groups. The immediate and follow-up complete occlusion rates for both groups were 100%. There was no device embolization, procedure-related stroke, or pericardial effusion.

CONCLUSIONS

The Ceraflex septal occluder is a safe and efficient device for closure of secundum atrial septal defects with no procedural complications. The Ceraflex has similar outcomes when compared with the Amplatzer septal occluder device. The advantage of the Ceraflex septal occluder device is that it can be deployed without the tension of the delivery catheter.

摘要

目的

Ceraflex房间隔缺损封堵器是Amplatzer封堵器的一种替代装置,具有一些结构创新,包括灵活连接、增加灵活性以及减少植入材料用量。我们评估了Ceraflex房间隔缺损封堵器经皮闭合继发孔型房间隔缺损的有效性和安全性。

患者与方法

这是一项前瞻性、非随机、多中心研究,纳入使用Ceraflex和Amplatzer房间隔缺损封堵器行经导管闭合房间隔缺损的患者。在1、6和12个月时进行临床评估及经胸超声心动图随访。

结果

2010年至2014年期间,125例患者在经食管超声心动图引导下使用Ceraflex房间隔缺损封堵器(n = 58)和Amplatzer房间隔缺损封堵器(n = 67)进行房间隔缺损闭合。两组患者的特征、缺损拉伸大小、封堵器大小及透视时间相似。两组的即刻及随访完全封堵率均为100%。未发生封堵器栓塞、与手术相关的卒中或心包积液。

结论

Ceraflex房间隔缺损封堵器是一种安全有效的继发孔型房间隔缺损闭合装置,无手术并发症。与Amplatzer房间隔缺损封堵器相比,Ceraflex有相似的结果。Ceraflex房间隔缺损封堵器的优点是可以在无输送导管张力的情况下展开。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验