Ruschke Stefan, Kienberger Hermine, Baum Thomas, Kooijman Hendrik, Settles Marcus, Haase Axel, Rychlik Michael, Rummeny Ernst J, Karampinos Dimitrios C
Department of Diagnostic and Interventional Radiology, Technische Universität München, Munich, Germany.
Bioanalytik Weihenstephan, Research Center for Nutrition and Food Sciences, Technische Universität München, Freising, Germany.
Magn Reson Med. 2016 Jan;75(1):32-41. doi: 10.1002/mrm.25578. Epub 2015 Mar 5.
To propose and optimize diffusion-weighted stimulated echo acquisition mode (DW-STEAM) for measuring fat unsaturation in the presence of a strong water signal by suppressing the water signal based on a shorter T2 and higher diffusivity of water relative to fat.
A parameter study for point-resolved spectroscopy (PRESS) and STEAM using oil phantoms was performed and correlated with gas chromatography (GC). Simulations of muscle tissue signal behavior using DW-STEAM and long-echo time (TE) PRESS and a parameter optimization for DW-STEAM were conducted. DW-STEAM and long-TE PRESS were applied in the gastrocnemius muscles of nine healthy subjects.
STEAM with TE and mixing time (TM) up to 45 ms exhibited R(2) correlations above 0.98 with GC and little T2 -weighting and J-modulation for the quantified olefinic/methylene peak ratio. The optimal parameters for muscle tissue using DW-STEAM were b-value = 1800 s/mm(2), TE = 33 ms, TM = 30 ms, and repetition time = 2300 ms. In vivo measured mean olefinic signal-to-noise ratios were 72 and 40, mean apparent olefinic water fractions were 0.19 and 0.11 for DW-STEAM and long-TE PRESS, respectively.
Optimized DW-STEAM MR spectroscopy is superior to long-TE PRESS for measuring fat unsaturation, if a strong water peak prevents the olefinic fat signal's quantification at shorter TEs and water's tissue specific ADC is substantially higher than fat.
提出并优化扩散加权刺激回波采集模式(DW-STEAM),通过基于水相对于脂肪更短的T2和更高的扩散率抑制水信号,在存在强水信号的情况下测量脂肪不饱和度。
使用油模对点分辨光谱法(PRESS)和STEAM进行参数研究,并与气相色谱法(GC)相关联。使用DW-STEAM和长回波时间(TE)PRESS对肌肉组织信号行为进行模拟,并对DW-STEAM进行参数优化。将DW-STEAM和长TE PRESS应用于9名健康受试者的腓肠肌。
TE和混合时间(TM)长达45 ms的STEAM与GC的R(2)相关性高于0.98,对于定量的烯烃/亚甲基峰比率几乎没有T2加权和J调制。使用DW-STEAM的肌肉组织的最佳参数为:b值 = 1800 s/mm(2),TE = 33 ms,TM = 30 ms,重复时间 = 2300 ms。在体内测量的平均烯烃信噪比分别为72和40,DW-STEAM和长TE PRESS的平均表观烯烃水分数分别为0.19和0.11。
如果强水峰在较短TE时阻止烯烃脂肪信号的定量,并且水的组织特异性表观扩散系数(ADC)明显高于脂肪,则优化的DW-STEAM磁共振波谱在测量脂肪不饱和度方面优于长TE PRESS。