Ventriculoatrial conduction: a cause of atrial malpacing in AV universal pacemakers. A report of two cases.

作者信息

van Gelder L M, El Gamal M I

出版信息

Pacing Clin Electrophysiol. 1985 Jan;8(1):140-3. doi: 10.1111/j.1540-8159.1985.tb05733.x.

Abstract

Retrograde atrial activation during ventricular pacing has often been a cause of intermittent or persistent arrhythmias (pacemaker-mediated tachycardia) in AV universal pacemakers. We recently encountered two cases in which VA conduction was responsible for atrial malpacing in patients with an implanted AV universal pacemaker, one programmed in DDD and one in DVI mode. Atrial malpacing was induced by the atrial refractoriness due to retrograde activation. In the first patient, it was observed when the pacemaker was programmed to a rate of 110 ppm (lower rate) and an AV interval of 200 ms in order to check crosstalk. In the second patient, it was observed after ventricular premature contractions.

摘要

在房室通用起搏器中,心室起搏期间的逆向心房激动常常是间歇性或持续性心律失常(起搏器介导的心动过速)的一个原因。我们最近遇到了两例,其中室房传导导致植入房室通用起搏器的患者出现心房起搏不良,一例起搏器程控为DDD模式,另一例为DVI模式。逆向激动导致的心房不应期诱发了心房起搏不良。在第一例患者中,当起搏器程控为110次/分钟(下限频率)且房室间期为200毫秒以检查串扰时观察到了这种情况。在第二例患者中,在室性早搏后观察到了这种情况。