Optimisation of paediatric chest radiography.

作者信息

Kostova-Lefterova D, Taseva D, Hristova-Popova J, Vassileva J

机构信息

National Centre of Radiobiology and Radiation Protection, Sofia, Bulgaria

National Centre of Radiobiology and Radiation Protection, Sofia, Bulgaria.

出版信息

Radiat Prot Dosimetry. 2015 Jul;165(1-4):231-4. doi: 10.1093/rpd/ncv119. Epub 2015 Apr 1.

Abstract

The purpose of this work is to present the algorithm for the optimisation of paediatric chest radiography aimed to reduce patient doses at keeping adequate image quality. Optimal tube voltage, tube current and exposure time, use of automatic or manual exposure control and use of antiscatter grid were recommended for different age groups and depending on the patient size. The optimised protocols and radiography technique resulted in decrease in Entrance surface air kerma and effective dose values in a factor of between 1.5 and over 5 for different age groups. Image quality was assessed to be of sufficient diagnostic quality.

摘要

这项工作的目的是提出一种优化儿科胸部X线摄影的算法,旨在在保持足够图像质量的同时降低患者剂量。针对不同年龄组并根据患者体型,推荐了最佳管电压、管电流和曝光时间、自动或手动曝光控制的使用以及防散射格栅的使用。优化后的方案和摄影技术使不同年龄组的体表空气比释动能和有效剂量值降低了1.5至5倍以上。经评估,图像质量具有足够的诊断质量。