¹¹¹铟-喷替肽定量SPECT/CT对正常腹部器官放射性示踪剂摄取的可重复性

Repeatability of Radiotracer Uptake in Normal Abdominal Organs with ¹¹¹In-Pentetreotide Quantitative SPECT/CT.

作者信息

Rowe Steven P, Vicente Esther, Anizan Nadège, Wang Hao, Leal Jeffrey P, Lodge Martin A, Frey Eric C, Wahl Richard L

机构信息

The Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland.

The Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland

出版信息

J Nucl Med. 2015 Jul;56(7):985-8. doi: 10.2967/jnumed.115.155358. Epub 2015 May 14.

Abstract

UNLABELLED

With an increasing emphasis on quantitation of SPECT imaging and its use in dosimetry to guide therapies, it is desirable to understand the repeatability in normal-organ SPECT uptake values (SPECT-UVs). We investigated the variability of normal abdominal organ uptake in repeated (111)In-pentetreotide SPECT studies.

METHODS

Nine patients with multiple (111)In-pentetreotide SPECT/CT studies for clinical purposes were evaluated. Volumes of interest were drawn for the abdominal organs and applied to SPECT-UVs. The variability of those values was assessed.

RESULTS

The average SPECT-UV for the liver (1.7 ± 0.6) was much lower than for the kidneys (right, 8.0 ± 2.4; left, 7.5 ± 1.7). Interpatient and intrapatient variability was similar (intraclass correlation coefficients, 0.40-0.59) for all organs. The average coefficients of variation for each organ for each patient were obtained and averaged across all patients (0.26 for liver, 0.22 for right kidney, and 0.20 for left kidney). The coefficients of variation for the organs across all scans were 0.33 (liver), 0.30 (right kidney), and 0.22 (left kidney).

CONCLUSION

Variability across all patients and all scans for the liver was higher than reported with (18)F-FDG PET, though left kidney variability was similar to PET liver variability and left kidney uptake may be able to serve as an internal metric for determining the quantifiability of an (111)In-pentetreotide SPECT study.

摘要

未标记

随着对单光子发射计算机断层扫描(SPECT)成像定量分析及其在剂量测定中用于指导治疗的重视程度不断提高,了解正常器官SPECT摄取值(SPECT-UVs)的可重复性是很有必要的。我们研究了重复进行的铟-111喷替膦SPECT研究中正常腹部器官摄取的变异性。

方法

评估了9例因临床目的进行多次铟-111喷替膦SPECT/CT研究的患者。为腹部器官绘制感兴趣区,并应用于SPECT-UVs。评估这些值的变异性。

结果

肝脏的平均SPECT-UV(1.7±0.6)远低于肾脏(右侧,8.0±2.4;左侧,7.5±1.7)。所有器官的患者间和患者内变异性相似(组内相关系数,0.40 - 0.59)。获取每位患者每个器官的平均变异系数,并在所有患者中进行平均(肝脏为0.26,右肾为0.22,左肾为0.20)。所有扫描中各器官的变异系数分别为:肝脏0.33、右肾0.30、左肾0.22。

结论

所有患者和所有扫描中肝脏的变异性高于氟-18氟代脱氧葡萄糖(18F-FDG)正电子发射断层扫描(PET)报告的变异性,尽管左肾变异性与PET肝脏变异性相似,且左肾摄取可能能够作为确定铟-111喷替膦SPECT研究可量化性的内部指标。

相似文献

引用本文的文献

本文引用的文献

9
Quantitative SPECT reconstruction using CT-derived corrections.使用CT衍生校正的定量SPECT重建。
Phys Med Biol. 2008 Jun 21;53(12):3099-112. doi: 10.1088/0031-9155/53/12/002. Epub 2008 May 21.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验