Department of Chemistry, Razi University, Kermanshah, Iran.
Department of Chemistry, Payame Noor University (PNU), PO Box 19395-3697, Tehran, Iran.
Mater Sci Eng C Mater Biol Appl. 2015 Nov 1;56:426-31. doi: 10.1016/j.msec.2015.07.001. Epub 2015 Jul 7.
Herein, the poly(o-phenylenediamine) decorated with gold-silver nanoparticle (Ag-Au NPs) nanocomposite modified glassy carbon was used for the determination of hydrogen peroxide. Electrochemical experiments indicated that the proposed sensor possesses an excellent sensitivity toward the reduction of hydrogen peroxide. The resulting sensor exhibited a good response to hydrogen peroxide over linear range from 0.2 to 60.0μM with a limit of detection of 0.08μM, good reproducibility, long-term stability and negligible interference from ascorbic acid, uric acid and dopamine. The proposed sensor was successfully applied to the determination of hydrogen peroxide in human serum sample.
在此,使用聚邻苯二胺修饰的金-银纳米粒子(Ag-Au NPs)纳米复合材料修饰的玻碳电极用于测定过氧化氢。电化学实验表明,该传感器对过氧化氢的还原具有优异的灵敏度。所得传感器对过氧化氢具有良好的响应,线性范围为 0.2 至 60.0μM,检测限为 0.08μM,具有良好的重现性、长期稳定性和对尿酸、抗坏血酸和多巴胺的可忽略干扰。该传感器成功应用于人血清样品中过氧化氢的测定。