Shah Anand D, Hirsh David S, Langberg Jonathan J
Division of Cardiology, Emory University School of Medicine, Atlanta, Georgia.
Pacing Clin Electrophysiol. 2016 Jan;39(1):101-4. doi: 10.1111/pace.12738. Epub 2015 Sep 17.
Defibrillator lead malfunction can be a disastrous complication, leading to loss of protection from sudden cardiac death in a high-risk patient population. Recognition of lead-specific risk for failure can assist in development of focused screening or surveillance, as in the case of the Riata lead (St. Jude Medical, St. Paul, MN, USA) or the Sprint Fidelis lead (Medtronic Inc., Minneapolis, MN, USA). A case of defibrillation failure secondary to a Durata lead insulation failure is presented. A brief review of the literature and current St. Jude Medical implantable cardiac defibrillator lead design is presented. Identification of arcing is identified as a potential sign of catastrophic insulation failure.
除颤器导线故障可能是一种灾难性的并发症,会导致高危患者群体失去预防心源性猝死的保护。识别特定导线的故障风险有助于开展针对性的筛查或监测,如Riata导线(美国明尼苏达州圣保罗市圣犹达医疗公司)或Sprint Fidelis导线(美国明尼苏达州明尼阿波利斯市美敦力公司)的情况。本文介绍了一例因Durata导线绝缘故障继发的除颤失败病例。还简要回顾了相关文献以及圣犹达医疗公司目前的植入式心脏除颤器导线设计。电弧的识别被确定为灾难性绝缘故障的一个潜在迹象。