Raposo Francisco, Borja Rafael, Cacho Jesús A, Mumme Jan, Mohedano Ángel F, Battimelli Audrey, Bolzonella David, Schuit Anthony D, Noguerol-Arias Joan, Frigon Jean-Claude, Peñuela Gustavo A, Muehlenberg Jana, Sambusiti Cecilia
Instituto de la Grasa, Consejo Superior de Investigaciones Científicas (CSIC), 41013 Seville, Spain.
Instituto de la Grasa, Consejo Superior de Investigaciones Científicas (CSIC), 41013 Seville, Spain.
J Chromatogr A. 2015 Sep 25;1413:94-106. doi: 10.1016/j.chroma.2015.08.008. Epub 2015 Aug 6.
The performance parameters of volatile fatty acids (VFAs) measurements were assessed for the first time by a multi-laboratory validation study among 13 laboratories. Two chromatographic techniques (GC and HPLC) and two quantification methods such as external and internal standard (ESTD/ISTD) were combined in three different methodologies GC/ESTD, HPLC/ESTD and GC/ISTD. Linearity evaluation of the calibration functions in a wide concentration range (10-1000mg/L) was carried out using different statistical parameters for the goodness of fit. Both chromatographic techniques were considered similarly accurate. The use of GC/ISTD, despite showing similar analytical performance to the other methodologies, can be considered useful for the harmonization of VFAs analytical methodology taking into account the normalization of slope values used for the calculation of VFAs concentrations. Acceptance criteria for VFAs performance parameters of the multi-laboratory validation study should be established as follows: (1) instrument precision (RSDINST≤1.5%); (2) linearity (R(2)≥0.998; RSDSENSITIVITY≤4%; REMAX≤8%; REAVER≤ 3%); (3) precision (RSD≤1.5%); (4) trueness (recovery of 97-103%); (5) LOD (≤3mg/L); and (6) LOQ (10mg/L).
13个实验室通过一项多实验室验证研究首次评估了挥发性脂肪酸(VFA)测量的性能参数。两种色谱技术(气相色谱法和高效液相色谱法)以及两种定量方法(外标法和内标法)被组合成三种不同的方法:气相色谱法/外标法、高效液相色谱法/外标法和气相色谱法/内标法。使用不同的统计参数对拟合优度进行评估,在宽浓度范围(10 - 1000mg/L)内对校准函数进行线性评估。两种色谱技术的准确性被认为相当。气相色谱法/内标法尽管与其他方法表现出相似的分析性能,但考虑到用于计算VFA浓度的斜率值的归一化,可被认为有助于VFA分析方法的统一。多实验室验证研究中VFA性能参数的验收标准应如下确定:(1)仪器精密度(RSDINST≤1.5%);(2)线性(R(2)≥0.998;RSDSENSITIVITY≤4%;REMAX≤8%;REAVER≤3%);(3)精密度(RSD≤1.5%);(4)准确度(回收率为97 - 103%);(5)检测限(≤3mg/L);以及(6)定量限(10mg/L)。