Paar Steffen, Umathum Reiner, Jiang Xiaoming, Majer Charles L, Peter Jörg
J Biomed Opt. 2015 Sep;20(9):095007. doi: 10.1117/1.JBO.20.9.095007.
A noncontact optical detector for in vivo imaging has been developed that is compatible with magnetic resonance imaging (MRI). The optical detector employs microlens arrays and might be classified as a plenoptic camera. As a resulting of its design, the detector possesses a slim thickness and is self-shielding against radio frequency (RF) pulses. For experimental investigation, a total of six optical detectors were arranged in a cylindrical fashion, with the imaged object positioned in the center of this assembly. A purposely designed RF volume resonator coil has been developed and is incorporated within the optical imaging system. The whole assembly was placed into the bore of a 1.5 T patient-sized MRI scanner. Simple-geometry phantom studies were performed to assess compatibility and performance characteristics regarding both optical and MR imaging systems. A bimodal ex vivo nude mouse measurement was conducted. From the MRI data, the subject surface was extracted. Optical images were projected on this surface by means of an inverse mapping algorithm. Simultaneous measurements did not reveal influences from the magnetic field and RF pulses onto optical detector performance (spatial resolution, sensitivity). No significant influence of the optical imaging system onto MRI performance was detectable.
已开发出一种用于体内成像的非接触式光学探测器,它与磁共振成像(MRI)兼容。该光学探测器采用微透镜阵列,可归类为全光相机。由于其设计,该探测器具有超薄的厚度,并且对射频(RF)脉冲具有自屏蔽功能。为了进行实验研究,总共六个光学探测器以圆柱形方式排列,成像对象位于该组件的中心。已开发出一种专门设计的射频体谐振器线圈,并将其纳入光学成像系统。整个组件被放置在一台1.5T患者尺寸的MRI扫描仪的孔内。进行了简单几何形状的体模研究,以评估光学和MR成像系统的兼容性和性能特征。进行了双模态离体裸鼠测量。从MRI数据中提取了受试者表面。通过逆映射算法将光学图像投影到该表面上。同时测量未发现磁场和RF脉冲对光学探测器性能(空间分辨率、灵敏度)有影响。未检测到光学成像系统对MRI性能有显著影响。