Zhao Shouqin, Gong Shusheng, Han Demin, Zhang Hua, Ma Xiaobo, Li Yi, Chen Xueqing, Ren Ran, Li Yuling
a Department of Otolaryngology Head and Neck Surgery , Beijing Tongren Hospital, Capital Medical University, Key Laboratory of Otolaryngology Head and Neck Surgery, Ministry of Education , Beijing , China ;
b Beijing Institute of Otolaryngology , Beijing , China.
Acta Otolaryngol. 2016;136(1):23-33. doi: 10.3109/00016489.2014.1003091. Epub 2015 Oct 23.
Application of the Vibrant Soundbridge to the round window (RW) membrane can be utilized as an efficient therapy for congenital oval window (OW) atresia.
To report the surgical technique and auditory outcome of an active middle ear implant (AMEI) system used in patients with congenital OW atresia.
Nine subjects with congenital OW atresia (six males and three females, ranging in age from 5.5 to 25 years, average 12.5 years) were implanted with an AMEI (Vibrant Soundbridge) at the round window (RW-Vibroplasty). Five cases were diagnosed as having isolated congenital OW atresia while four patients presented with combined external/middle ear malformation.
An improvement of 30 dB in average pure-tone air conduction thresholds (0.5-4 kHz) was achieved, with the high frequencies showing greater results. The subjects achieved postoperative speech recognition scores of 80-100% on the Computerized Mandarin Speech Test System (CMSTS) sentence test. Bone conduction thresholds were confirmed as stable in all subjects postoperatively. Decline in auditory benefit was noticed in two subjects, who then underwent revision surgery. One of these revision surgery patients then experienced stable hearing recovery, while the other patient's hearing declined.
将振动声桥应用于圆窗膜可作为先天性卵圆窗闭锁的有效治疗方法。
报告用于先天性卵圆窗闭锁患者的有源中耳植入(AMEI)系统的手术技术及听觉结果。
9例先天性卵圆窗闭锁患者(6例男性,3例女性,年龄5.5至25岁,平均12.5岁)接受了圆窗(RW - 振动成形术)的AMEI(振动声桥)植入。5例被诊断为孤立性先天性卵圆窗闭锁,4例患者伴有外耳道/中耳联合畸形。
平均纯音气导阈值(0.5 - 4 kHz)提高了30 dB,高频改善更为明显。患者在计算机普通话言语测试系统(CMSTS)句子测试中的术后言语识别率达到80 - 100%。术后所有受试者的骨导阈值均保持稳定。2例患者出现听觉益处下降,随后接受了翻修手术。其中1例翻修手术患者听力恢复稳定,而另1例患者听力下降。