Schick Daniel, Pratap Jit
1 Biomedical Technology Services, Health Support Queensland, Queensland, Australia.
2 Division of Diagnostic Radiology, Princess Alexandra Hospital, Brisbane, Australia.
Br J Radiol. 2016;89(1058):20150486. doi: 10.1259/bjr.20150486. Epub 2015 Nov 11.
This study evaluated the radiation dose and image quality implications of dual-energy CT (DECT) use, compared with kilovoltage-optimized single-source/single-energy CT (SECT) on a dual-source Siemens Somatom(®) Definition Flash CT scanner (Siemens Healthcare, Forcheim, Germany).
With equalized radiation dose (volumetric CT dose index), image noise (standard deviation of CT number) and signal-difference-to-noise ratio (SDNR) were measured and compared across three techniques: 100, 120 and 100/140 kVp (dual energy). Noise in a 30-cm-diameter water phantom and SDNR within unenhanced soft-tissue regions of a small adult (50 kg/165 cm) anthropomorphic phantom were utilized for the assessment.
Water phantom image noise decreased with DECT compared with the lower noise SECT setting of 120 kVp (p = 0.046). A decrease in SDNR within the anthropomorphic phantom was demonstrated at 120 kVp compared with the SECT kilovoltage-optimized setting of 100 kVp (p = 0.001). A further decrease in SDNR was observed for the DECT technique when compared with 120 kVp (p = 0.01).
On the Siemens Somatom Definition Flash system (Siemens Healthcare), and for equalized radiation dose conditions, image quality expressed as SDNR of unenhanced soft tissue may be compromised for DECT when compared with kilovoltage-optimized SECT, particularly for smaller patients.
DECT on a dual-source CT scanner may require a radiation dose increase to maintain unenhanced soft-tissue contrast detectability, particularly for smaller patients.
本研究评估了在双源西门子Somatom(®) Definition Flash CT扫描仪(德国福希海姆西门子医疗公司)上,与千伏优化单源/单能量CT(SECT)相比,双能量CT(DECT)的辐射剂量和图像质量影响。
在三种技术条件下(100、120和100/140 kVp(双能量)),在辐射剂量均衡(容积CT剂量指数)的情况下,测量并比较图像噪声(CT值标准差)和信号差异噪声比(SDNR)。使用直径30厘米的水模中的噪声以及小体型成人(50千克/165厘米)仿真人体模型未增强软组织区域内的SDNR进行评估。
与120 kVp的较低噪声SECT设置相比,DECT扫描的水模图像噪声降低(p = 0.046)。与SECT千伏优化设置100 kVp相比,120 kVp时仿真人体模型内的SDNR降低(p = 0.001)。与120 kVp相比,DECT技术的SDNR进一步降低(p = 0.