Accuracy of linear measurements using cone beam computed tomography and panoramic radiography in dental implant treatment planning.

作者信息

Luangchana Penporn, Pornprasertsuk-Damrongsri Suchaya, Kiattavorncharoen Sirichai, Jirajariyavej Bundhit

出版信息

Int J Oral Maxillofac Implants. 2015 Nov-Dec;30(6):1287-94. doi: 10.11607/jomi.4073.

Abstract

PURPOSE

The aim of this study was to investigate the accuracy of linear measurements from cone beam computed tomography (CBCT) images and digital panoramic radiographs at various implant sites.

MATERIALS AND METHODS

Fifty implant sites from six skulls were marked with gutta-percha and subjected to CBCT with five different voxel protocols: 0.125 mm, 0.160 mm, and 0.250 mm with the 3D Accuitomo 170 CBCT machine and 0.200 mm and 0.300 mm with the CS 9500 CBCT machine. Images were also taken with the CS 9000 panoramic machine with three protocols: normal head, chin-up, and chin-down positions. Electronic linear measurement of bone height using the corresponding machine's software was recorded by two observers. Physical measurement using a digital caliper with ± 0.02-mm accuracy was directly recorded at the corresponding regions as the gold standard. All image measurements were compared with the physical measurements. The paired sample correlations for physical measurement, mean difference, standard deviation, absolute error, absolute percentage error, and inter- and intraobserver reliability were calculated.

RESULTS

Intraobserver and interobserver reliability was more than 0.99. Paired sample correlation between all image measurements and physical measurements was considered statistically significant at P < .05. All image measurements were underestimated by less than 2 mm, except for the chin-down position of the maxilla in the panoramic radiograph. The absolute error and absolute percentage error in the mandible were less than those in the maxilla, and values obtained with CBCT were less than those from panoramic radiographs.

CONCLUSION

CBCT images using the 3D Accuitomo 170 and CS 9500 machines and digital panoramic radiographs via a picture archiving and communication system are sufficiently accurate for vertical linear measurements in dental implant treatment planning.

摘要

目的

本研究旨在调查锥形束计算机断层扫描(CBCT)图像和数字化全景X线片在不同种植部位进行线性测量的准确性。

材料与方法

在六个颅骨上的五十个种植部位用牙胶进行标记,并使用五种不同体素协议进行CBCT扫描:使用3D Accuitomo 170 CBCT机的0.125毫米、0.160毫米和0.250毫米,以及使用CS 9500 CBCT机的0.200毫米和0.300毫米。还使用CS 9000全景机在三种协议下进行图像采集:正常头部、抬头和低头位置。两名观察者使用相应机器的软件对骨高度进行电子线性测量。使用精度为±0.02毫米的数字卡尺在相应区域直接进行物理测量作为金标准。将所有图像测量结果与物理测量结果进行比较。计算物理测量的配对样本相关性、平均差异、标准差、绝对误差、绝对百分比误差以及观察者间和观察者内的可靠性。

结果

观察者内和观察者间的可靠性均超过0.99。所有图像测量结果与物理测量结果之间的配对样本相关性在P <.05时被认为具有统计学意义。除全景X线片中上颌的低头位置外,所有图像测量结果的低估均小于2毫米。下颌的绝对误差和绝对百分比误差小于上颌,并且CBCT获得的值小于全景X线片获得的值。

结论

使用3D Accuitomo 170和CS 9500机器的CBCT图像以及通过图像存档和通信系统的数字化全景X线片对于牙种植治疗计划中的垂直线性测量具有足够的准确性。