Houbart Virginie, Cobraiville Gaël, Servais Anne-Catherine, Napp Aurore, Merville Marie-Paule, Fillet Marianne
Laboratory for the Analysis of Medicines, Department of Pharmacy, CIRM, University of Liege, Belgium.
Department of Clinical Chemistry, CHU, University of Liege, Belgium.
Bioanalysis. 2015 Nov;7(21):2789-99. doi: 10.4155/bio.15.181. Epub 2015 Nov 13.
Dried blood analysis experiences a growing interest due to practical, ethical and financial advantages compared with classical wet plasma or serum analysis. Besides classical DBS, new alternatives are commercialized as volumetric absorptive microsampling (VAMS) that are expected to overcome hematocrit influence.
The feasibility of hepcidin (a peptide hormone) extraction and determination from DBS and VAMS blood sampling was investigated. Experimental design was used to determine the optimal extraction conditions. Matrix effect and extraction recovery were studied and a special attention was paid to phospholipid removal.
The data suggest that the combination of VAMS and phospholipid removal plates provides low matrix effect and high sensitivity, and constitutes an easy and promising protocol for hepcidin analysis.
与传统的湿血浆或血清分析相比,干血分析因其实际、伦理和经济优势而越来越受到关注。除了传统的干血斑(DBS)外,新的替代方法如体积吸收微采样(VAMS)已商业化,有望克服血细胞比容的影响。
研究了从DBS和VAMS血样中提取和测定铁调素(一种肽激素)的可行性。采用实验设计确定最佳提取条件。研究了基质效应和提取回收率,并特别关注磷脂的去除。
数据表明,VAMS与磷脂去除板相结合可提供低基质效应和高灵敏度,是一种简便且有前景的铁调素分析方法。