Yamamoto Yohei, Yamaguchi Satoshi, Sasho Takahisa, Fukawa Taisuke, Akatsu Yorikazu, Nagashima Kengo, Takahashi Kazuhisa
Department of Orthopedic Surgery, Graduate School of Medical and Pharmaceutical Sciences, Chiba University, Chiba, Japan (Y.Y., S.Y., T.S., T.F., Y.A., K.T.); and Clinical Research Center, Chiba University Hospital, Chiba, Japan (K.N.).
J Ultrasound Med. 2016 Jan;35(1):159-66. doi: 10.7863/ultra.14.11042. Epub 2015 Dec 11.
The purposes of this study were to measure intraobserver and interobserver repeatability of quantitative elastography using an acoustic coupler for the Achilles tendon, to compare elastographic values among different age groups, and to assess the correlation between quantitative and conventional qualitative measurements.
One hundred asymptomatic Achilles tendons of 50 volunteers were examined. For quantitative elastography, the strain value of the tendon was divided by that of the acoustic coupler with a known Young modulus, and the strain ratio was calculated. B-mode image assessment and qualitative elastography were also performed. Intraobserver repeatability and interobserver repeatability of strain ratio measurements were calculated. The strain ratios were compared among age groups by the Kruskal-Wallis test. Additionally, strain ratios for each B-mode grade and qualitative elastographic grade were compared by the Wilcoxon signed rank test.
Intraobserver repeatability intraclass correlation coefficient (1, 3) values were 0.87 and 0.93, respectively. The correlation coefficient between the observers' measurements was 0.61. The strain ratio for the 30s age group was 0.27, which was significantly lower than the values for the other age groups (P < .001). Although 97 of 100 tendons were normal on B-mode evaluations, 36 of the 97 tendons were degenerated on qualitative elastography and also had higher strain ratio values than normal tendons.
Quantitative elastography using an acoustic coupler is a reproducible technique for measuring the elasticity of the Achilles tendon. It may detect early tendon degeneration that is not depicted on B-mode imaging.
本研究旨在测量使用声学耦合器对跟腱进行定量弹性成像时观察者内和观察者间的重复性,比较不同年龄组的弹性成像值,并评估定量测量与传统定性测量之间的相关性。
对50名志愿者的100条无症状跟腱进行检查。对于定量弹性成像,将肌腱的应变值除以具有已知杨氏模量的声学耦合器的应变值,计算应变比。还进行了B模式图像评估和定性弹性成像。计算应变比测量的观察者内重复性和观察者间重复性。通过Kruskal-Wallis检验比较各年龄组的应变比。此外,通过Wilcoxon符号秩检验比较每个B模式等级和定性弹性成像等级的应变比。
观察者内重复性组内相关系数(1,3)值分别为0.87和0.93。观察者测量值之间的相关系数为0.61。30岁年龄组的应变比为0.27,显著低于其他年龄组的值(P <.001)。虽然100条肌腱中有97条在B模式评估中正常,但97条肌腱中有36条在定性弹性成像中退变,且应变比值也高于正常肌腱。
使用声学耦合器的定量弹性成像是一种可重复的测量跟腱弹性的技术。它可能检测到B模式成像未显示的早期肌腱退变。