Direct epicardial assist device using artificial rubber muscle in a swine model of pediatric dilated cardiomyopathy.

作者信息

Saito Yoshiaki, Suzuki Yasuyuki, Kondo Norihiro, Kowatari Ryosuke, Daitoku Kazuyuki, Minakawa Masahito, Fukuda Ikuo

机构信息

Department of Thoracic and Cardiovascular Surgery, Hirosaki University School of Medicine, Aomori - Japan.

出版信息

Int J Artif Organs. 2015 Nov;38(11):588-94. doi: 10.5301/ijao.5000447. Epub 2015 Dec 7.

Abstract

PURPOSE

Ventricular assist devices are a potent alternative or bridge therapy to heart transplants for dilated cardiomyopathy patients. However, ventricular assist devices have problems related to biocompatibility, hemocompatibility, and thromboembolic events, especially in younger patients. The present study examined the hemodynamic effects of a direct cardiac compression device using circumferential artificial rubber muscles in a young swine model of dilated cardiomyopathy.

METHODS

Dilated cardiomyopathy was established in 6 pigs (6-8 weeks of rapid right ventricular pacing; average weight, 22.6 ± 2.1 kg). The device was designed using pneumatic rubber muscles (Fluidic Muscle, Festo). Hemodynamic parameters were monitored under baseline conditions, after the assistance, and after inducing ventricular fibrillation. Hemodynamic data were acquired using a PiCCO, multilumened thermodilution catheter in the pulmonary artery, left ventricular pressure monitoring, and epicardial echocardiography.

RESULTS

Direct epicardial assistance resulted in a significant improvement in hemodynamic data. Cardiac output improved from 1.39 ± 0.24 L/min to 1.96 ± 0.46 (p = 0.02). Stroke volume (14.5 ± 3.2 mL versus 20.1 ± 4.3 ml, p<0.01) and ejection fraction (25.2 ± 3.6% versus 47.7 ± 7.8%, p<0.01) also improved after assistance. After inducing ventricular fibrillation, cardiac output was maintained at 1.33 ± 0.28 L/min.

CONCLUSIONS

Use of a circumferential direct epicardial assistant device resulted in improvement in hemodynamic data in a dilated cardiomyopathy model. Although there is still a need for improvements in device components, the direct cardiac assist device may be a good alternative to recent heart failure device therapies.

摘要

目的

对于扩张型心肌病患者,心室辅助装置是心脏移植的有效替代或过渡治疗方法。然而,心室辅助装置存在与生物相容性、血液相容性和血栓栓塞事件相关的问题,尤其是在年轻患者中。本研究在年轻的扩张型心肌病猪模型中,研究了使用环形人造橡胶肌肉的直接心脏压迫装置的血流动力学效应。

方法

对6头猪(6 - 8周龄,快速右心室起搏;平均体重22.6±2.1千克)建立扩张型心肌病模型。该装置采用气动橡胶肌肉(流体肌肉,费斯托公司)设计。在基线条件下、辅助后以及诱发心室颤动后监测血流动力学参数。使用脉搏指示连续心输出量监测仪(PiCCO)、肺动脉内的多腔热稀释导管、左心室压力监测和心外膜超声心动图获取血流动力学数据。

结果

直接心外膜辅助使血流动力学数据有显著改善。心输出量从1.39±0.24升/分钟提高到1.96±0.46升/分钟(p = 0.02)。辅助后每搏输出量(14.5±3.2毫升对20.1±4.3毫升,p<0.01)和射血分数(25.2±3.6%对47.7±7.8%,p<0.01)也有所改善。诱发心室颤动后,心输出量维持在1.33±0.28升/分钟。

结论

在扩张型心肌病模型中,使用环形直接心外膜辅助装置可改善血流动力学数据。尽管该装置组件仍需改进,但直接心脏辅助装置可能是近期心力衰竭装置治疗的良好替代方案。